mA Output (P200 to P219)

When an mA Output Parameter is accessed, the mA symbol is displayed in the Index Type
field and the mA output number (corresponding to the SITRANS LU01 terminals) is
displayed in the Index field.
P200 mA Output Range
Determines the mA output range.
Primary Index
Values
Related
If either 1 or 2 is selected, the mA output is directly proportional to the mA Function. If
either 3 or 4 is selected, then the output is inversely proportional.
P201 mA Output Function
Alters the mA output/measurement relationship.
Primary Index
Values
Related
Altered By
P203 mA Output Value / Transducer
Displays current mA output value.
Values
This displays as an Auxiliary Reading when
does not include adjustments made using Trim features (P214 / P215).
This parameter is applicable only if any mA output has the transducer
Note:
Index Number as its input source (see P201).
7ML19985BE02
mA output
0
off
1
0 to 20 mA
2
*
4 to 20 mA
3
20 to 0 mA
4
20 to 4 mA
P911 mA Output Value
mA output
value
mA function
0
OFF
1
level
2
space
3
distance
4
volume
P202 mA Output Allocation
P911 mA Output Value
P001 Operation
Range: 0.000 to 22.00 (view only)
SITRANS LU01 – INSTRUCTION MANUAL
20 mA = high level (unless P001 = 2
or 3, then 20 mA = low)
20 mA = low level (unless P001 = 2
or 3, then 20 mA = high)
Operation (P001)
level
space
distance
key is pressed in the RUN mode and
Page 47