3. Specifications
3.1. Analog Section
Input sensitivity
Input voltage range
Zero adjustment range
Load cell excitation voltage
Load cell drive capacity
Temperature
Zero
coefficient
Span
Nonlinearity
A/D conversion method
A/D resolution count
Display resolution
Sampling rate
3.2. Digital Section
Display element
Measurement display
Status indicators
Key switches
0.15 μV/d or greater
-35 mV to +35 mV (-7 mV/V to +7 mV/V)
-35 mV to +35 mV (-7 mV/V to +7 mV/V)
5VDC 5%, 120 mA with remote sensing capability
Maximum 8 x 350 Ω load cells
0.02 μV /°C typ., 0.1 μV /°C max.
3ppm /°C typ., 15ppm /°C max.
0.005 % of full scale
Delta-sigma method
Approximately 16,000,000 counts
999,999 d max.
(Less than 20,000 d recommended)
100 times/second
• Measurement display: 6-digit 7-segment green LED
• Character height: 14.6 mm
• Polarity display: 1 green LED
• Status indicators: 6 red LEDs
• Switches between NET and GROSS
• Selectable decimal places (10
• Overflow display
All the digits turn OFF. (When the polarity is negative, the
minus sign appears at the highest-order digit.)
ZERO, STABLE, GROSS, NET, HOLD,
ZERO, TARE, NET/GROSS, F, ON/OFF, ENTER, CAL
Page 5
1
2
3
4
, 10
, 10
, 10
, 10
5
)
AD-4410