A&D AD-4410 Instruction Manual page 79

Weighing indicator
Table of Contents

Advertisement

Function No.
Parameter or
Setting range
To judge when the negative gross weight is
exceeded.
Calf13
A/D negative exceeded or,
1
3
to
1: Gross weight < -999999
2: Gross weight < negative weighing capacity
3: Gross weight < -19 d
To judge when the negative net weight is
exceeded.
Calf14
Gross weight negative exceeded or,
1
2
to
1: Net weight < -999999
2: Net weight < negative weighing capacity
Select whether or not to clear the zero value.
Calf15
0: Disables.
0
1
to
1: Enables.
Select whether or not to perform zero upon
Calf16
power-ON.
0
1
0: Disables.
to
1: Enables.
Calf17
Input voltage from a load cell at zero.
-7. 0 0000
to
X.XXXXX mV/V
7. 0 0000
Calf18
Input voltage from a load cell at span
0. 0 0001
to
X.XXXXX mV/V
9. 9 9999
Calf19
The
1
999999
corresponding to the span input voltage
to
Select whether or not to output when the
Calf20
gross weight is negative.
0
1
0: Enables output.
to
1: Disables output.
Select whether or not to restrict
Calf21
network-related communication.
0
1
0: Disables restriction.
to
1: Enables restriction.
Second header for serial output
Calf22
1: GS / NT / TR
1
2
to
2: G_ / N_ / T_ ( _ is space, ASCII 20h)
Number of unit digits for serial output
Calf23
1: 2 digits
1
2
to
2: 3 digits
Scale interval of dual range
1: 1
2: 2
Calf24
3: 5
1
6
to
4: 10
5: 20
6: 50
Description
value
of
the
Page 77
calibration
weight
Default
User
value
setting
1
1
1
0
0. 0 0000
3. 2 0000
32000
0
0
1
1
1
AD-4410