10.2.4. Checking the RS-232C

While
Chcr51L
transmitted using the baud rate set in general functions.
When the data is received, it is displayed for two seconds.
Test data
S T , G S , + 0 0 0 0 0 . 0 k g

10.2.5. Checking the RS-232C / RS-485 (Option)

While
Chcr52L
transmitted using the baud rate set in general functions.
When the data is received, it is displayed for two seconds.
Test data
S T , G S , + 0 0 0 0 0 . 0 k g

10.2.6. Checking the control I/O

While Chc io is displayed, press the ENTER key to enter control I/O mode. In control
I/O mode, the control I/O state is displayed as ON (
Pressing the TARE,
output 3, respectively.
Output 3 Output 2 Output 1 Input 3
ENTER

10.2.7. Checking the A/D (load cell)

With
displayed, press the ENTER key to enter the A/D check mode. The load
Chc ad
cell output value is displayed in mV/V.
The example below is for an internal count of 1.23456 mV/V.
1 . 2 3 4 5 6
If the value exceeds ±7mV/V, it may be because of mechanical damage or a connection
error involving the load cell. Refer to "10.4. Confirming the load cell connection with a
digital multi meter" to check the connection.

10.2.8. Checking the internal count

With
displayed, press the ENTER key to enter the internal count check mode.
Chc in
The internal count value (display value times 10) is displayed.
The example below is for an internal count of 123.
is displayed, each time the ENTER key is pressed, test data is
is displayed, each time the ENTER key is pressed, test data is
F
or ENTER key will return the state of output 1, output 2 or
TARE
The decimal point blinks
1 2 3 .
C R L F
C R L F
) or OFF (
Input 2 Input 1
The decimal point blinks
Page 63
) on the display.
AD-4410