9.2.5. Bit structure

Data 7 bits
Data 8 bits

9.2.6. Data transfer mode

Four data transfer modes are available; stream, auto printing, manual printing or jet
stream.
Stream mode
Auto printing mode
Manual printing mode
Jet stream
Functions related to the data transfer mode
Cl f04
Set
,
output.
Example:
@
0 9
ID number
Calf20
Set
weight is negative.
Calf22
Set
AD-4410
Data bit
Start bit
Transmits data in sync with the display update rate. If
the indicator can not catch up with the update rate
because of the baud rate, the transmission is paused
until the next updating.
The transmitted data is the same as what is being
displayed. Data that is not displayed is never
transmitted.
Transmits data one time once a weight value of 5 d or
greater has stabilized.
For a further transmission, the weight value must fall
below 5 d and then stabilize again above that value.
Transmits data when the F key, assigned as the
manual print command, is pressed.
Transmits data in sync with the sampling rate. If the
indicator can not catch up with the sampling because
of the baud rate, the transmission is paused until the
next sampling.
r51f06
r52f06
or
S T , G S , + 0 1 2 3 . 4 5 k g
(disables negative output) to disable the print command when gross
(header 2) to change header 2 to G, N or T.
Data bit
(ID number) to add an ID number to standard serial
Page 54
Stop bit
Parity bit
Stop bit