Hold - A&D AD-4410 Instruction Manual

Weighing indicator
Table of Contents

Advertisement

8.4. Hold

Function No.
Parameter
range
LHldf01
0. 0 0
9. 9 9
to
LHldf02
0. 0 0
9. 9 9
to
LHldf03
0
2
to
LHldf04
0
1
to
LHldf05
0. 0 0
9. 9 9
to
LHldf06
0
999999
to
LHldf07
0
1
to
* This hold function only works when
hold). This hold function has no function when
Hldf01
*
(averaging time) works only when
Function
Time to perform the averaging.
Averaging
Units in seconds.
time
0.00 is not averaged.
Time to wait before commencing hold or
Start wait
averaging.
time
Units in seconds.
Condition to start the hold or averaging
Condition of
automatically
0: Do not use automatic start.
automatic
start
1 Above the near-zero range, and stable
2 Above the near-zero range
Release due
Release when control input is falling
to falling
0: Do not release
control input
1: Release
Release after a set amount of time
has passed
Time
Units in seconds.
release
Release will not happen if the value is set
to 0.00.
Release when fluctuation from the
holding value exceeds a set value
Release due
The decimal point position is linked to
to range of
Calf02
fluctuation
Release will not happen if the value is set
to 0.
Release when the weighing value is in
Release in
the near-zero range
near-zero
0: Do not release
range
1: Release
Description
.
fncf07
(hold) is set to
fncf07
fncf07
(hold) is set to
Page 45
2
(peak hold) or
1
(hold) is set to
(normal hold).
3
(averaging hold).
Default
value
0. 0 0
0. 0 0
0
1
0. 0 0
0
0
3
(averaging