Actual load setting
Set digital linearization by loading/unloading masses.
* Warm up the instrument for at least ten minutes to avoid the effects of temperature drift.
* The input order should proceed from the smallest mass to the largest mass.
Step 1 Remove the calibration switch cover on the lower
left of the front panel, and press the CAL key (the
switch inside).
the instrument can go into calibration mode.
Press the ENTER key to enter into calibration
mode, at which point Cal5et is displayed. Press
the
or
ENTER key.
Step 2
lnr 0 is displayed. If you want to monitor the
current weighing value, press the CAL key. The
gross weight is displayed.
If the CAL key is pressed again,
displayed.
Step 3 With nothing placed on the load cell, wait for the
stabilization indicator to turn ON and then press
the ENTER key
seconds.
lnr 1 is displayed. Press the ENTER key. The
Step 4
mass value is displayed (current weight value
setting), and the last digit of the mass value
blinks.
Press the < > ∧ ∨ keys to adjust the number to
the mass that should be inputted.
If you want to check the current weighing value,
press the CAL key. The gross weight is displayed.
If the CAL key is pressed again, the mass value is
displayed.
Step 5 Place a calibration weight on the load cell. Wait
for the stabilization indicator to turn ON and then
press the ENTER key. ------ is displayed for
two seconds.
Cal
is displayed, meaning
key to select lnr5et , then press the
------ is displayed for two
.
lnr 0
Page 29