Emerson Rosemount 3300 Series Reference Manual page 41

Guided wave radar level and interface transmitters
Hide thumbs Also See for Rosemount 3300 Series:
Table of Contents

Advertisement

Reference Manual
00809-0100-4811, Rev CA
February 2006
Coaxial
L>49 inches/
1250 mm
Maximum shortening
23.6 inches / 600 mm
Minimum probe length
15.7 inches / 400 mm
L≤49 inches/
1250 mm
Rosemount 3300 Series
To cut a coaxial probe do the
following:
Centering piece
NOTE!
The HTHP coaxial probe must not be
cut in field.
1. Insert the centering piece.
2. Cut the tube to the desired length.
3. Move the centering piece.
4. Cut the rod inside the tube. Make
• Pipes longer than
• Pipes shorter than
5. Update the transmitter
(The centering piece is
delivered from factory and
should be used to prevent the
spacers centering the rod from
coming loose).
sure that the rod is fixed with the
centering piece while cutting.
49 inches/1250 mm can be
shortened by as much as
23.6 inches/600 mm.
49 inches/1250 mm can be cut as
long as the remaining length is not
less than
15.7 inches/400 mm.
configuration to the new probe
length, see "Probe Length" on
page 4-9.
3-17