Brady Tagus User Manual page 83

Table of Contents

Advertisement

Step
Operator Action
7
Proceed according to your media
type:
Die-cut stock: strip the media from
the backing and then reinsert it into
the sensor; adjust the Sensor Eye
Mark over the center of the backing.
Notched stock: adjust the Sensor
Eye Mark over the center of the
notch.
Continuous stock: go to
Step 8.
Press the
Note: Do NOT move the position of the Media Sensor after it has been adjusted.
8
Position the label (and backing, if
any) under the Sensor Eye Mark.
Note: If using preprinted media,
ensure the label area under the
sensor is free of preprinted text,
graphics or borders.
Press the
9
Observe the display.
Quick Test Mode
SCAN BACKING
PRESS ESC KEY <YYY>
key to continue.
SCAN PAPER
PRESS ESC KEY <YYY>
key to continue.
CALIBRATION COMPLETE
Or, for reflective media:
CALIBRATION COMPLETE
Or, for continuous media:
CALIBRATION COMPLETE
Displayed Message
GAP MODE
REFLECTIVE MODE
CONTINUOUS MODE
Comment
This sets the parameter for the
'gap' or 'mark' value.
Where 'yyy' is a numerical
value representing the current
sensor reading: useful in
locating the best
sensor position.
Note: Never position the
sensor over a perforation
when recording a sensor
reading.
This sets the parameter for the
'paper' value.
Where 'yyy' is a numerical
value representing the current
sensor reading.
The calibration was successful.
See note below if another
message was displayed.
5-37 Rev A