Reading Rate (Auto-Zero Off)
Temperature Coefficient (Auto-Zero off)
For a stable environment ±1°C add the
following additional error for AZERO OFF
Range
Error
100 mV–10 V
5 µV/°C
100 V–1000 V
500 µV/°C
2 / Resistance
Two-wire and Four-wire Ohms (OHM and OHMF Functions)
Range
Full Scale
Maximum
Resolution
10 W
12.00000
10 µW
100 W
120.00000
10 µW
1 kW
1.2000000
100 µW
10 kW
12.000000
1 mW
100 kW
120.00000
10 mW
1 MW
1.2000000
100 mW
1 W
10 MW
12.000000
9
10 W
100 MW
120.00000
7
1 GW
1.2000000
100 W
Selected Reading Rates
NPLC Aperture Digits Bits A-Zero
0.0001 1.4 µs
0.0006 10 µs
2
0.01
167 µs
0.1
1.67 ms
1
16.6 ms
10
0.166 s
100
1000
Maximum Input
HI to LO
4
LO to Guard
Guard to Earth
Input Terminals
Terminal Material: Gold-plated Tellurium Copper
Input Leakage Current:<20pA at 25°C
Test
Open
Current
Voltage
Circuit
6
Source
10 mA
0.1 V
12 V
1 mA
0.1 V
12 V
1 mA
1.0 V
12 V
100 µA
1.0 V
12 V
50 µA
5.0 V
12 V
5 µA
5.0 V
12 V
500 nA
5.0 V
12 V
500 nA
5.0 V
5 V
500 nA
5.0 V
5 V
1
Readings / Sec
Off
3
4.5
16
100,000
4,130
5.5
18
50,000
3,150
6.5
21
5,300
930
2
6.5
21
592
245
2
7.5
25
60
29.4
2
8.5
28
6
3
8.5
28
36/min
18/min
8.5
28
3.6/min
1.8/min
Rated Input Non-Destructive
±1000 V pk ±1200 V pk
±200 V pk
±350 V pk
5
±500 V pk
±1000 V pk
Maximum
Maximum
Lead
Series Offset
Resistance
(OCOMP
(OHMF)
ON)
20 W
0.01 V
200 W
0.01 V
150 W
0.1 V
1.5 kW
0.1 V
1.5 kW
0.5 V
1.5 kW
1.5 kW
1.5 kW
1.5 kW
Appendix A Specifications
1.
For PRESET; DELAY 0;
DlSP OFF; OFORMAT
DINT; ARANGE OFF.
A-Zero
2.
Aperture is selected
On
independent of line
frequency (LFREQ).
These apertures are for 60
Hz NPLC values where
1 NPLC = 1/LFREQ. For
50 Hz and NPLC
indicated, aperture will
increase by 1.2 and
reading rates will decrease
by 0.833
3.
For OFORMAT SINT
10
W LO to Guard
4.
> 10
with guard open.
12
W Guard to Earth.
5.
> 10
Temperature Coefficient (ppm
of Reading + ppm of Range) / °C
Without
With ACAL
7
ACAL
3+1
1+1
3+1
1+1
3+0.1
1+0.1
3+0.1
1+0.1
3+0.1
1+0.1
3+1
1+1
20+20
5+2
100+20
25+2
1000+20
250+2
6.
Current source is ± 3%
absolute accuracy.
7.
Additional error from Tcal or
last ACAL ± 1° C.
8.
Additional error from Tcal ±
5° C.
9.
Measurement is computed
from 10 M W in parallel with
input
8