Download Print this page

Agilent Technologies 3458A Calibration Manual page 91

Hide thumbs Also See for 3458A:

Advertisement

5 / AC Current
AC Current (ACI and ACDCI Functions)
Maximum
Range
Full Scale
Resolution
100 µA
120.0000
100 pA
1 mA
1.200000
1 nA
10 mA
12.00000
10 nA
100 mA
120.0000
100 nA
1 A
1.050000
1 µA
2
AC Accuracy
24 Hour to 2 Year (% Reading + % Range)
10 Hz to
Range
20 Hz
4
0.4+0.03
100 µA
1 mA – 100 mA 0.4+0.02
1 A
0.4+0.02
AC + DC Accuracy (ACDCI Function)
For ACDCI Accuracy apply the following additional error to the ACI accuracy.
(% of Reading + % of Range).
DC£10% of AC
Accuracy
Temperature Coefficient
0.005+0.02
0.0+.001
Additional Errors
Apply the following additional errors as appropriate to your particular measurement setup.
LOW Frequency Error ( % of Reading )
ACBAND Low
Signal
10 Hz-1 kHz
1 to 10 kHz
Frequency
NPLC >10
NPLC >1
10–200 Hz
0
200–500 Hz 0
0.15
500–1 kHz
0
0.015
1–2 kHz
0
0
2–5 kHz
0
0
5–10 kHz
0
0
6
Reading Rates
ACBAND Low
NPLC
³ 10 Hz
10
³1 kHz
1
³10 kHz
0.1
Shunt
Burden
Resistance
Voltage
730 W
0.1 V
100 W
0.1 V
10 W
0.1 V
1 W
0.25 V
0.1 W
< 1.5 V
20 Hz to
45 Hz to
100 Hz to
45 Hz
100 Hz
5 kHz
0.15+0.03 0.06+0.03
0.06+0.03
0.15+0.02 0.06+0.02
0.03+0.02
0.16+0.02 0.08+0.02
0.1+0.02
DC>10% of AC
5
Accuracy
0.15+0.25
>10 kHz
NPLC >0.1
0.9
0.2
0.05
0.01
Maximum Sec / Reading
ACI
ACDCI
1.2
1
1
0.1
1
0.02
Temperature Coefficient
(% of Reading + % of Range) / °C
0.002+0
0.002+0
0.002+0
0.002+0
0.002+0
5 kHz to
20 kHz to
3
20 kHz
50 kHz
0.06+0.02
0.4 +0.04
0.3+0.02
1+0.04
Temperature Coefficient
0.0+0.007
Crest Factor Error (% of Reading)
Crest Factor
Additional Error
1 – 2
0
2 – 3
0.15
3 – 4
0.25
4 – 5
0.40
1.
Additional error beyond
±1°C, but within ±5°C of last
1
ACAL.
2.
Specifications apply full
scale to 1/20 full scale, for
sine wave inputs, crest factor
= 1.4, and following
PRESET within 24 hours
and ±1°C of last ACAL.
Add 5 ppm of reading
additonal error for Agilent
factory traceabiltiy to US
NIST. Traceability is the
sum of the 10V and 10 k
50 kHz to
traceability values.
3
3
100 kHz
3.
Typical performance
4.
1 kHz maximum on the 100
0.55+0.15
µA range.
5
5.
Additional error beyond
±1°C, but within ±5°C of
last ACAL
(% of Reading + % of
Range) /° C.
6.
For DELAY–1; ARANGE
OFF. For DELAY 0;
NPLC.1, unspecified reading
rates of greater than
500/sec are possible.
Appendix A Specifications
W
91

Hide quick links:

Advertisement

loading