Agilent Technologies 3458A Calibration Manual page 32

Hide thumbs Also See for 3458A:

Advertisement

Adjusted by:
3458A serial number or other device ID number:
Previous calibration number (CALNUM? command):
multimeter)
Adjustments performed:*
1.
Front Terminal Offset Adjustment
2.
Rear Terminal Offset Adjustment
3.
DC Gain Adjustment (DCV Standard Uncertainty =
4.
Resistance and DC Current Adjustment (Resistance Standard Uncertainty =
5.
AC Adjustment:
Multimeter
Reading
Line A
Line B
Line C
Line D
Line E
Line F
Line G
Line H
Internal adjustment temperature (TEMP? command):
Calibration number (CALNUM? command):
Calibration secured:
*Always perform the above adjustments numbered 1 through 4: adjustment number 5 is only required once every
2 years or whenever the 03458-66502 or 03458-66503 PC Assembly has been replace or repaired.
32 Chapter 3 Adjustment Procedures
3458A Adjustment Record
Synthesizer
AC source
setting
frequency
V
3V
3V
V
V
V
1V
V
V
100mV
V
unsecured:
Date:
(record this number before adjusting the
Execute
Command
100 kHz
SCAL 1E5
100 kHz
SCAL 10
2 MHz
SCAL 10
8 MHz
SCAL 10
100 kHz
SCAL 1
8 MHz
SCAL 1
100 kHz
SCAL .1
8 MHz
SCAL .1
°
C
(record this number after adjusting the multimeter)
)
)
Adjustment Description
Frequency Adjustment
Low-freq. voltage reference
Time interpolator & flatness
Flatness Adjustment
Low-freq voltage reference
Flatness Adjustment
Low-freq voltage reference
Flatness Adjustment