Dynamic Performance
100 mV, 1 V, 10 V Ranges; Aperture = 6 µs
Test
DFT-harmonics
DFT-spurious
Differential non-linearity
Signal to Noise Ratio
Direct and Sub-sampled Digitizing (DSDC, DSAC, SSDC and SSAC Functions)
Input
Range 1
Impedance
10 mV
1 MW with 140 pF <50 µV
100 mV
1 MW with 140 pF <90 µV
1 V
1 MW with 140 pF <800 µV
10 V
1 MW with 140 pF <8 mV
100 V
1 MW with 140 pF <80 mV
1000 V
1 MW with 140 pF <800 mV
DC to 20 kHz Performance
2
0.02 % of Reading + Offset
Maximum Sample Rate
Function
Readings / sec
SSDC, SSAC
100 M (effective)
DSDC, DSAC 50 k
Dynamic Performance
100 mV, 1 V, 10 V Ranges; 50,000 Samples/sec
Test
Input (2 x full scale pk–pk) Result
DFT-harmonics
20 kHz
DFT-harmonics
1.005 MHz
DFT-spurious
20 kHz
Differential non-linearity 20 kHz
Signal to Noise Ratio
20 kHz
94
Appendix A Specifications
Input (2 x full scale pk-pk) Result
1 kHz
1 kHz
dc
1 kHz
Offset
Typical
2
Bandwidth
Voltage
2 MHz
12 MHz
12 MHz
12 MHz
12 MHz
3
2 MHz
Resolution
4
16 bits
16 bits
< –96 dB
< –100 dB
< 0.003% of Range
>96 dB
3
Sample Timebase
Accuracy: 0.01 %
Jitter: < 100 ps rms
<–90 dB
<–60 dB
External Trigger
<–90 dB
Latency: < 125 ns
<0.005 % of Range
Jitter: < 2 ns rms
>66 dB
Level Trigger
Latency: < 700 ns
Jitter: < 100 ps, for 1 MHz
full scale input
1.
Maximum DC voltage limited
to
400 V DC in DSAC or SSAC
functions.
2.
±1°C and within 24 hours of
last ACAL ACV.
3.
Limited to 1 x10
product.
4.
Effective sample rate is
determined by the smallest
time increment used during
synchronous sub-sampling of
the repetitive input signal,
which is 10 ns.
5.
<25 ns variability between
multiple 3458As