CALNUM?
Description
Remarks
Example
Calibration Number Query. Returns a decimal number indicating the number
of times the multimeter has been adjusted.
Syntax
CALNUM?
•
The calibration number is incremented by 1 whenever the multimeter is
unsecured and adjusted. If autocal is secured, the calibration number is
also incremented by 1 whenever an autocal is performed; if unsecured,
autocal does not affect the calibration number.
•
The calibration number is stored in cal-protected memory and is not lost
when power is removed.
•
The multimeter was adjusted before it left the factory which increments
the CALNUM. When you receive the multimeter, read the calibration
number to determine its initial value.
•
Related Commands: CAL. CALSTR. SCAL
10 OUTPUT 722;"CALNUM?" !READS CALIBRATION NUMBER
20 ENTER 722;A
30 PRINT A
40 END
!ENTERS RESPONSE INTO COMPUTER
!PRINTS RESPONSE
Chapter 5 Command Summary
73