standard).
Note
Voltage standards from 1V DC to 12V DC can be used for this procedure.
However, using a voltage standard <10V DC will degrade the multimeter's
accuracy specifications.
1.
Select the DC Voltage function.
2.
Set the front panel Terminals switch to Front.
3.
Connect the voltage standard to the multimeter's front panel HI and
LO Input terminals as shown in Figure 2. If using a Guard wire (as
shown in Figure 2). set the Guard switch to the Open position. If not
using a Guard wire, set the Guard switch to the To LO position.
4.
Execute the CAL command specifying the exact output voltage of the
standard. For example, if the standard's voltage is 10.0001 VDC,
execute CAL 10.0001. The multimeter automatically performs the DC
gain adjustment and the display shows each of the various steps being
performed. This adjustment takes about 2 minutes. When the
adjustment is complete, the multimeter returns to displaying DC
voltage measurements.
5.
Disconnect the voltage standard from the multimeter.
Figure 2. DC Gain Adjustment Connections
Chapter 3 Adjustment Procedures
25