7 / Digitizing Specifications
General Information
The 3458A supports three independent methods for signal digitizing. Each method is discussed below to aid in
selecting the appropriate setup best suited to your specific application.
DCV
Standard DCV function.
This mode of digitizing allows signal acquisition at rates from 0.2 readings / sec at
28 bits resolution to 100k readings / sec at 16 bits. Arbitrary sample apertures from
500 ns to 1 sec are selectable with 100 ns resolution. Input voltage ranges cover
100 mV to 1000 V full scale. Input bandwidth varies from 30 kHz to 150 kHz
depending on the measurement range.
DSDC
Direct Sampling DC Coupled measurement technique.
DSAC
Direct Sampling AC Coupled measurement technique.
In these modes the input is sampled through a track / hold with a fixed 2 ns aperture
which yields a 16 bit resolution result. The sample rate is selectable from 6000 sec
/ sample to 20 µs / sample with 100 ns resolution. Input voltage ranges cover 10
mV peak to 1000 V peak full scale. The input bandwidth is limited to 12 MHz.
SSDC
Sub-Sampling ( Effective time sampling ) DC Coupled.
SSAC
Sub-Sampling ( Effective time sampling ) AC Coupled.
These techniques implement synchronous sub-sampling of a repetitive input signal
through a track / hold with a 2 ns sample aperture which yields a 16 bit resolution
result. The effective sample rate is settable from 6000 sec / sample to 10 ns /
sample with 10 ns resolution. Sampled data can be time ordered by the instrument
and output to the GPIB. Input voltage ranges cover 10 mV peak to 1000 V peak full
scale. The input bandwidth is limited to 12 MHz.
Summary of Digitizing Capabilities
Technique
Function
Standard
DCV
Direct-sampled DSDC / DSAC
Sub-sampled
SSDC / SSAC
Standard DC Volts Digitizing (DCV Function)
Offset
Input
Range
Impedance
Voltage
100 mV
10
<5 µV
W
>10
1 V
10
<5 µV
W
>10
10 V
10
<5 µV
W
>10
100 V
10 MW
<500 µV
1000 V
10 MW
<500 µV
DC Performance
0.005% of Reading + Offset
Maximum Sample Rate (See DCV for more data)
Readings / sec
Resolution
100 k
15 bits
100 k
16 bits
50 k
18 bits
Input Bandwidth Best Accuracy
DC – 150 kHz
0.00005 – 0.01%
DC – 12 MHz
0.02%
DC – 12 MHz
0.02%
Typical
Settling Time
1
Bandwidth
to 0.01% of Step
80 kHz
50 µs
150 kHz
20 µs
150 kHz
20 µs
30 kHz
200 µs
30 kHz
200 µs
1
Aperture
0.8 µs
1.4 µs
6.0 µs
Sample Rate
100 k/sec
50 k/sec
100 M / sec (effective)
Sample Timebase
Accuracy: 0.01 %
Jitter: < 100 ps rms
External Trigger
2
Latency: < 175 ns
Jitter: < 50 ns rms
Level Trigger
Latency: < 700 ns
Jitter: < 50 ns rms
Appendix A Specifications
1.
±1° C of an AZERO or
within
24 hours and ±1°C of last
ACAL.
2.
<125 ns variability between
multiple 3458As
93

Hide quick links:

Advertisement

loading