Settling Characteristics
For first reading or range change error using default delays, add .01% of input step additional error for the 100
µA to 100 mA ranges. For the 1 A range add .05% of input step additional error.
The following data applies for DELAY 0.
Function
ACBAND Low
³10 Hz
ACI
ACDCI
10 Hz – 1 kHz
1 kHz – 10 kHz
³10 kHz
Maximum Input
Rated lnput
I to LO
± 1.5 A pk
LO to Guard
± 200 V pk
Guard to Earth
± 500 V pk
6 / Frequency/ Period
Frequency / Period Characteristics
Voltage (AC or DC Coupled)
ACV or ACDCV Functions
Frequency Range
1 Hz – 10 MHz
Period Range
1 sec – 100 ns
Input Signal Range
700 V rms – 1 mV rms
Input Impedance
1 MW±15% with<140 pF
Accuracy
24 Hour- 2 Year
Range
0°C-55°C
1 Hz–40 Hz
1 s–25 ms
0.05% ofReading
40 Hz – 10 MHz
25 ms–100 ns
.01% ofReading
Measurement Technique:
Reciprocal Counting
Time Base:
10 MHz ± 0.01%, 0°C to 55°C
Level Trigger:
±500% of Range in 5% steps
92
Appendix A Specifications
DC Component
Settling Time
DC < 10% AC
0.5 sec to 0.01%
DC > 10% AC
0.9 sec to 0.01%
0.5 sec to 0.01%
0.08 sec to 0.01%
0.015 sec to 0.01%
Non-Destructive
< 1.25A rms
± 350 V pk
± 1000 V pk
Current (AC or DC Coupled)
1
ACI or ACDCI Functions
1 Hz – 100 kHz
1sec – 10 µs
1 A rms – 10 µA rms
0.1 – 730 W
Reading Rates
Resolution
Gate Time
0.00001%
1 s
>0.0001%
100 ms
> 0.001%
10 ms
> 0.01%
1 ms
> 0.1%
100 µs
Trigger Filter:
Selectable 75 kHz Low Pass Trigger Filter
Slope Trigger:
Positive or Negative
1
2
3
4
Readings/sec
0.95
9.6
73
215
270
1.
The source of frequency
measurements and the
measurement input
coupling are determined by
the FSOURCE command.
2.
Range dependent, see ACI
for specific range
impedance values.
3.
Gate Time is determined
by the specified
measurement resolution.
4.
For Maximum Input
specified to fixed range
operation. For auto range,
the maximum speed is 30
readings/sec for ACBAND
³ 1 kHz.
Actual Reading Speed is
the longer of 1 period of
the input, the chosen gate
time, or the default reading
time-out of
1.2 sec.