RPM4-AD™ OPERATION AND MAINTENANCE MANUAL
4.4.4

PROGRAM MESSAGE DESCRIPTIONS

#
Purpose
Classic
Arguments
Remarks
Example
(classic)
See Also
ABORT
Purpose
Command
Classic
Remarks
Example
(enhanced)
Example
(classic)
See Also
ARANGEn
Purpose
Errors
ATM
Purpose
Query
Classic
Remarks
Example
(enhanced)
Example
(classic)
Errors
See Also
AUTOZEROn
Purpose
Command
Query
Classic
Default
Optional Suffix
Arguments
© 2005-2007 DH Instruments, a Fluke Company
To allow the host PC to communicate with a device connected to the RPM4-AD COM2
port.
"#xx"
xx:
The string to send out of the RPM4-AD COM2 port. It must be less than
40 characters long.
The RPM4-AD COM2 port can be used to communicate to another RS232 device (such
as another RPM4). This allows the user to use a single COM port or IEEE-488 port on
the host computer to communicate with the RPM4-AD and another RS232 device. A
carriage return and a line feed (<CR><LF>) are added to the string.
After this program message is issued, the RPM4-AD will reply back every string received
by the RPM4-AD COM2 port that is terminated with a carriage return. Line feeds are
discarded. This will discontinue when the next query is sent to the RPM4-AD.
There is no other reply from this program message. Prior to using this program message,
you must ensure that the RPM4-AD COM2 port is correctly set up to communicate with
the device on COM2. Refer to the "COM2=" program message.
Sent:
"#VER"
Reply:
"DH INSTRUMENTS, INC RPM4-AD us A1000/A0015 Ver2.00 "
This example assumes that a second RPM4's COM1 port is connected to the RPM4-AD
COM2 port. This example gets the version of the second RPM4-AD.
"COM2"
3.5.2
Cancels any pending data requests by the "PR", "PRR" "RATE", "SR" or "CONT"
commands.
"ABORT"
"ABORT"
This program message is usually used to ensure the RPM4-AD is in an idle condition.
Cmd sent:
"ABORT"
Reply:
"ABORT" (no reply if IEEE-488)
Sent:
"ABORT"
Reply:
"ABORT"
"PR", "PRR" "RATE", "SR" or "CONT"
Not used in RPM4-AD models
ERR# 53:
Command not available.
Reads the next measured pressure from RPM4's on-board barometer (if present).
"ATM?"
"ATM"
The atmospheric pressure as measured by the RPM4-AD on-board barometer is returned
in the current pressure units of the active Q-RPT (always absolute). This measurement is
followed by the units text.
Query sent:
"ATM?"
Query reply:
"97.12348 kPa a"
Sent:
"ATM"
Reply:
"97.12384 kPa a"
ERR# 23:
RPM4-AD is not equipped with a barometer.
3.5.7.1
Read or set the ON/OFF status of the AutoZ function.
"AUTOZEROn state"
"AUTOZEROn?"
"AUTOZEROn= state"
"AUTOZEROn"
"AUTOZERO1"
"n"
The active Q-RPT is assumed if no suffix is given.
'1'
Specify the Hi Q-RPT if the HL RPT is not active
Specify the HL Q-RPT if the HL Q-RPT is active
'2'
Specify the Lo Q-RPT if the HL Q-RPT is not active
'3'
Specify the HL Q-RPT if the HL Q-RPT is active
'state':
'0' Autozero OFF
'1' Autozero ON
Page 88