3.5.5.3
<3TIME>
PURPOSE
To view and edit the RPM4-AD internal time and date settings.
OPERATION
To
access
the
TIME
press [SPECIAL], <5prefs>, <3time>.
The display is:
Select <1time> to edit the time.
pressing [ENT] after each entry. Seconds go to zero when minutes are entered.
This can be used to synchronize the time with a time standard.
Select <2date> to edit the date. The date must be specified in YYYYMMDD format.
The RPM4-AD date and time are set to United States Mountain Standard
Time in the final test and inspection process at the factory. If desired, use
the TIME and DATE function to set your local time and date.
3.5.5.4
<4ID>
PURPOSE
To view or edit the RPM4-AD user ID and to view the RPM4-AD serial number.
OPERATION
To access the ID function press [SPECIAL] and select <5prefs>, <2ID>.
Select <1view> to view the current ID.
Select <2edit> to edit the ID.
The ID has twelve characters. When the edit screen is opened, the cursor is on
the first character. Numerical values can be entered directly from the keypad. In
addition, the cursor control keys can be used to toggle through a list of available
alphanumeric characters.
Character order going up is: blank space, symbols, lower case letters, upper
case letters, numbers. After selecting a character, press [ENT] to activate it and
move to the next character field.
When a character is selected the cursor moves to the next character. To leave a
blank character, press [ENT] with the field for that character blank. Use this for
the trailing characters if the ID being entered is less than twelve characters.
After the last of the twelve characters has been entered, the <Save ID?> option is offered.
Select <1no> to return to the ID edit screen. Select <2yes> to save the edited ID.
The ID can be set remotely from a computer which is quite a bit more
convenient than entering characters from the keyboard (see Section 4.4.4,
ID command). The ID is not cleared or reset by any RESET function (see
Section 3.5.9).
Page 67
function
Edit:
08:32:11 am 20030125
Edit hours, then minutes, then am/pm by
Holding the key slews through the characters.