Setup Menu (Acp) - Anritsu MS27200A Measurement Manual

Spectrum analyzer, remote spectrum monitor, spectrum monitor module
Table of Contents

Advertisement

Spectrum Analyzer Measurements

SETUP Menu (ACP)

The Adjacent Channel Power SETUP menu is available in MEASURE > MEASUREMENT > ACP > SETUP.
Once the ACP measurement is selected, the SETUP menu can be quickly accessed by tapping on the summary
display area below the spectrum window.
Figure 3-75. SETUP Menu (ACP)
Spectrum Analyzer MG
CHANNEL SPACING: Sets the channel frequency spacing (center channel to center
channel).
MAIN INTEG BW: Sets the range of integration used in calculating the power within the
channel. The integration bandwidth (IBW) is displayed as two vertical dashed lines for
each channel.
ADJ INTEG BW: Sets the range of integration bandwidth used in calculating the power
in the adjacent channel.
ALT INTEG BW: Sets the range of integration used in calculating the power in the
alternate channel.
LIMIT TYPE: Selects Absolute or Relative for the limit evaluation.
LIMITS: Turns on or off the use of ACP limits.
MAIN CH LIMIT: Sets the main channel limit.
ADJ CH LIMIT: Sets the adjacent channel limit.
ALT CH LIMIT: Sets the alternate channel limit.
PRESET SETUP: Sets all ACP setup parameters to default. Turns off limits.
PN: 10580-00447 Rev. J
3-17 Adjacent Channel Power
3-91

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MS27200A and is the answer not in the manual?

Table of Contents