11.2
Disinfection
11.3
Maintenance
Instruction manual OCULUS Easygraph (G/70620/XXXX/EN - Rev02)
11 Cleaning, Disinfection and Maintenance
Cleaning the Placido Bowl
The Placido bowl is a precision component and is pressure-sensitive.
The surfaces of these components are susceptible to scratching.
Take special care when cleaning the surface of the Placido bowl.
Use a lint-free, dry cloth.
Make sure that no dust gets into the little holes.
If necessary, carefully clean the Placido bowl with a barely damp
cloth.
Attention
Risk of electric shock if the OCULUS Easygraph is not completely
disconnected from the mains for these jobs.
Turn the OCULUS Easygraph off,
Pull the power plug before cleaning. When disconnecting electrical
connections, pull on the respective plug and not on the cable itself.
Recommendation: Use disinfection wipes suitable for medical
devices., for example:
Mikrozid sensitive wipes premium
Fa. Schülke & Mayr
Softpack 48 pieces; Art.Nr. 165711
Schülke & Mayr GmbH
Telefon: +4940521000; Telefax: +494052100318
mail@schuelke.com
www.schuelke.com
Note
Equipment damage caused by disinfectant solution
The disinfectant solution may damage the surface of the device if it is
sprayed directly on it.
Spray the disinfectant solution onto a cleaning cloth, do not spray
it directly on the device.
Turn the OCULUS Easygraph off,
Disinfect the forehead rest after each examination, disinfect the
housing if necessary.
If you do not use chin rest paper, disinfect the forehead rest after
each patient.
The OCULUS Easygraph is designed so that no special maintenance is
necessary. For safety reasons, we recommend that the illumination and
electrical values be checked every two years.
Please contact OCULUS Service for this.
sec. 8.3, page
25.
sec. 8.3, page
25.
43 / 62