Page 3
In particular, pay attention to the safety instruc- tions. The operating concept of the Oculus Pentacam® and Pentacam® HR is basical- ly identical.
Patient group ..................... 21 6 Device Description ................22 Overview of the device components ............22 Applied parts...................... 23 Functionality of the Pentacam® / Pentacam® HR ......23 7 Setup and Installation................25 Setup ........................25 4 / 72 Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00)
Page 5
11.2.3 Cleaning the illuminated slit............51 11.3 Disinfection ......................51 11.4 Test measurements..................52 11.5 Maintenance....................... 52 11.6 Attaching Paper to the Chin Rest............... 53 12 Troubleshooting..................55 Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00) 5 / 72...
Page 6
Electromagnetic Emissions and Immunity for the Pentacam® / Pentacam® HR....................63 17.3 Description of the Connection ..............67 17.4 Data sheet HEMG 49-S240210-7 (05150150)........68 17.5 Instructions for integration into an IT-Network ........70 6 / 72 Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00)
Floating License Key 77900 with manual SI/77900/.../en Viewing License Pentacam® Fast Screening Report software modul 70927 Full Sequence Measurement 10006911 Pentacam® Data-USB-Stick 017090901001 Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00) 7 / 72...
Page 8
Have the damages noted on the bill of lading, so that your claim for damages can be handled properly. sec. 13, For more information regarding shipping and handling, see page 56 8 / 72 Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00)
UDI number (example), consisting of: Model number UDI-DI (Device-Identification) UDI-PI (Product Identifier) and machine-readable matrix code Safety class Article number Disposal in household trash is pro- Device type hibited Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00) 9 / 72...
On the packaging Symbol Description Keep dry This way up Fragile Transport Limit of temperature for transport Storage Limit of temperature for storage Limit of humidity Limit of air pressure 10 / 72 Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00)
3 Structure of the Documentation Structure of the Documentation A folder containing documentation is supplied with your Pentacam® / Pentacam® HR: Short Guide: This document describes the measurement process and can used as a checklist. This document supports you when...
Pentacam® / Pentacam® HR are given in the Instruction Manual for the unit. It is imperative that you read and understand the whole Instruction Manual before you use the Pentacam® / Pentacam® HR. Carefully read through the Instruction Manual.
Transport and Storage Instructions Refer to the notes in chap. “13 Transport and Storage” on page 56. Instructions for Setup and Connection Only OCULUS or an authorized dealer is allowed to set up and to connect the device. Do not use or store the device in rooms that are humid, see chap.
(ME equipment) or with another person being in contact with the ME equipment. Only use devices that are compliant with IEC 60601-1 in the patient environment. Fig. 4-1: Patient environment 14 / 72 Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00)
All devices of the ME system must comply with the requirements of IEC 60601-1 or IEC 62368-1. Instructions for Operation Before first use: Let OCULUS or an authorized dealer train you in the operation of the device. Never operate a damaged device.
225 treatments (per patient eye). Instructions for Maintenance In order to retain the high measurement accuracy of the device OCULUS Optikgeräte GmbH recommends to perform a maintenance service every two years or after 25000 scans. A corresponding message appears.
Recommendation: Maintain a minimum distance (refer "17.1 Electromagnetic Compatibility” on page 62). If the distance is shorter, you must ensure that the Pentacam® / Pentacam® HR functions correctly. 4.12 Instructions on Cyber Security The device does not require a network or internet connection to work.
4.12.5 Unavailable Service Users should report unavailable service or prohibited access to information to their local Healthcare Organization’s IT department. 18 / 72 Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00)
The usage of a firewall (software or hardware) is recommended. Follow the instructions for integration into an IT network (see chap. “17.5 Instructions for integration into an IT-Network” on page 70). Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00) 19 / 72...
cornal thickness. Intended Medical Indication The Pentacam® / Pentacam® HR is indicated as an aid to screen several eye diseases for example, but not limited to: Keratoconus classification and progression Early ectatic diseases ...
5 Intended Purpose Intended users The Pentacam® / Pentacam® HR is intended exclusively for use in: ophthalmology practices clinics opticians or optometrists The Pentacam® / Pentacam® HR is intended for use by trained personnel: Who can ensure proper handling based on their knowledge, ...
Scheimpflug images created during an examination are transmitted to the connected PC. Up to 50 (Pentacam) or 100 (Pentacam HR) Scheimpflug images can be captured within maximum two seconds. Up to 25,000 (HR: 138,000) genuine height values are measured and analysed from the Scheimpflug images.
Page 24
6 Device Description Caution OCULUS Optikgeräte GmbH shall not be liable in any form for further use of the data recorded by a Pentacam® and for the evaluations it has calculated. 24 / 72 Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00)
Handle the device with care. Set up the Pentacam® / Pentacam® HR so that the power plug is easy to access. That way, you can easily disconnect it from the power supply for any repairs or maintenance work.
Plug of the Y-cable Fig. 7-1: Electrical Connection Attention If the device is not connected correctly and voltage is applied, the device may be damaged after a short time. 26 / 72 Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00)
The Pentacam-Software is network compatible. This makes it possible to install the Pentacam® Software on several PCs, connected in a local network. Make sure, that all PC in a network do have the same Pentacam® software version installed. A Floating License Key is part of every Pentacam® shipment. Please ensure a proper installation based on the Manual for the Floating License Key.
After starting, the patient data management opens (see chap. “10 Patient Data Management” on page 42). The Pentacam® program is not to be used to dictate therapy without professional correlation with other clinical findings and diagnostic tests. ®...
Page 29
The Pupil Image (2) shows the vertical / horizontal positioning of the Pentacam. The object is to position the yellow dot marking the apex of the cornea in the centre of the crosshairs. The blue dot marks the centre of the pupil, while the blue ring frames the pupil.
You can also use the Return key. Inside the adjustment window (7), there are arrows showing you the direction in which you must move the Pentacam® to activate automatic measurement (Automatic Release). This line (8) shows messages about the device, for example if a ...
8.2.2 Parameters applicable only to the Pentacam® HR Note The parameters described below apply only to the Pentacam® HR. The Pentacam® HR has additional options for selecting image types (see Scan Menu “Image Options" box (3)). "Cornea Fine" group box Select this option for a more detailed image of the cornea.
The Pentacam® program will load the examination you have selected. Online help The Pentacam® program provides you a direct help. You can recognize that by a yellow mark. Examples: Open the direct help by clicking on the yellow mark.
Page 33
Examination mode Imag- Automatic meas- Notes examination urement IOLs, ICLs, PIOLs 3D scan for PIOLs, 25-50 If the pupil is insufficiently dilated, possibly longer ex- apply mydriatic drops posure Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00) 33 / 72...
Due to small movements of the patient or caused by a wheelchair, the patient is no longer positioned appropriately to the Pentacam® / Pentacam® HR, which can lead to incorrect measurements. Perform a Pentacam® / Pentacam® HR scan only if the patient sits in a stationary chair.
Example of a rough adjustment for the right eye: Move the cross slide (1) until the marking at the end of the cross slide (4) roughly coincides with the circle R (3) on the sliding plate. Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00) 35 / 72...
Page 36
Look at the patient’s eye you are examining from one side and make sure that the blue slit light illuminates the cornea. If necessary, adjust the position of the cross slide to the left or right. 36 / 72 Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00)
Page 37
Focus the pupil image by moving the joystick towards the Pentacam® or away from it. Adjust the left/right position of the Pentacam® and its height setting. Move the joystick to the left or right and rotate the joystick clockwise or anticlockwise.
Rotate the joystick counter-clockwise When you have achieved the expected position, a cross will appear in the centre of the ring, surrounded by four lines. The Pentacam® will automatically begin measuring, alternately you can start the measuring procedure manually.
9 Measuring Procedure Quality Specifications After you have begun measuring either automatically or manually, the Pentacam® program opens. The "QS" button appears in a field below the patient data. Fig. 9-5: Pentacam® program with "QS" button No. Description "QS" button All examinations are automatically saved, regardless of the quality of the measurement taken.
Page 40
CSP Fixation If this value exceeds the permissible threshold, you must repeat the measurement. If necessary explain to the patient that he has to fix the black ring. 40 / 72 Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00)
Page 41
Terminate the current examination which has been saved. If required, make preparations to examine another patient. In the "Pentacam" overview, select the "Examination" menu and click [New Patient/End]. Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00) 41 / 72...
[Export] button [Import] button Patient list Patient data group box To get into the Pentacam® / Pentacam® HR program, you must first enter a new patient (8) or select an existing patient from the examination list (2). 10.2 Entering a New Patient Press the [New] button to enter a new patient in the Patient Data ...
Page 44
10 Patient Data Management up a list of any previous examinations for that patient in the examination window (bottom right side). 44 / 72 Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00)
Click [Export] button below the patient list. The following dialog appears: Fig. 10-5: "Export patient data" dialog No. Description Saving destination selection [...] button [Cancel] and [Export] buttons Options for data export Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00) 45 / 72...
Page 46
You can send data stored on the hard drive as an e-mail attachment. Requirements for transferring data to another PC: The Pentacam® program must be installed on the other PC. If the program is updated on the Pentacam® PC (sender), the program on the other PC (recipient) must also be updated.
(e.g. selection of the directory). Select the option (1) where the source data are contained ("Folder" or "U12" (single file)). Recommendation: Export the patient data using the "Single file (U12)" option. Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00) 47 / 72...
10.8.1 Backup Data Press the [Backup] button at the top right of the Patient Data Management system. The following dialog is displayed: Fig. 10-7: "Backup" dialog 48 / 72 Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00)
Patient Data Management system. The settings required for this can be made in the "Settings" area, User Guide In the patient data management you record and manage patient data. Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00) 49 / 72...
Maintenance by OCULUS Service or an authorized dealer after 25 000 measurements 11.2 Cleaning Attention Risk of electric shock if the Pentacam® / Pentacam® HR is not completely disconnected from the mains for these jobs. sec. 7.5, page 27 Turn the Pentacam® / Pentacam® HR off, ...
Disinfect the forehead rest after each examination. If you do not use paper for the chinrest, disinfect the chinrest after each examination. Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00) 51 / 72...
11 Cleaning, Disinfection and Maintenance 11.4 Test measurements The Pentacam® / Pentacam® HR is tested and calibrated in the Oculus factory. OCULUS Optikgeraete GmbH recommends to perform regular test measurements of the Pentacam® / Pentacam® HR. A human eye should be measured initially. Minimum 5 consecutive measurements per eye should be done and the arithmetic mean has to be calculated and recorded.
Fig. 11-4: Information when service is due In examinations (it will be stored): Fig. 11-5: Sign to perform maintenance Let the Pentacam® / Pentacam® HR checked by our service depart- ment or an authorized dealer. 11.6 Attaching Paper to the Chin Rest If you want to attach new chin rest paper, follow these instructions: Fig.
Page 54
11 Cleaning, Disinfection and Maintenance Insert the two pins (1) in the chin rest. 54 / 72 Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00)
Pentacam® on as soon as Patient Data Management becomes active. When you start the Pentacam® program, the message, "Load Bootloader" must appear. Contact the service department or your authorised dealer Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00) 55 / 72...
13 Transport and Storage Transport and Storage The Pentacam® / Pentacam® HR, must be properly dismantled and packed before being transported or stored. Caution Risk of equipment damage due to incorrect shipment or from improper storage Avoid shocks, vibrations, and contamination.
Loosen the screw connection of the Y cable and pull it out. When disconnecting electrical connections, pull on the respective plug and not on the cable itself. Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00) 57 / 72...
Assumption of Liability for Functions and Damage OCULUS will only accept responsibility for the safety, reliability and serviceability of the Pentacam® / Pentacam® HR if the unit is used in compliance with the following terms: Only use the equipment in conformance with this instruction ...
Power adapter HEMG49-S240210-7 (05150150) Mains connection 100 – 240 V AC Frequency 50/60 Hz Power input, max. 85 VA Output voltage 24 V DC Fuses Integrated overcurrent shut-off 60 / 72 Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00)
CE in accordance with Regulation (EU) 2017/745 on Medical Devices (MDR) The unit is a class IIa product. Conformity assessment procedure: (EU) 2017/745 MDR, Annex IX, chap- ter I and III. Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00) 61 / 72...
Caution The use of accessories, transducers and cables specified by OCULUS with devices other then the Pentacam® / Pentacam® HR may result in increased emissions or decreased immunity of the other device. 62 / 72 Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00)
Guidance and manufacturer´s declaration electromagentic emmissions IEC 60601-1-2 The OCULUS Pentacam® / Pentacam® HR is intended for operation in the electromagnetic environment specified below. The user of the Pentacam® / Pentacam® HR should ensure that it is being used in such an environment. Emissions test...
Page 64
0% U ; 250/300 0% U ; 250/300 τ τ periods periods Note: U is the a.c. mains voltage prior to application of the test level τ 64 / 72 Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00)
Page 65
To assess the electromagnetic environment due to fixed RF transmitters, and electromagnetic site survey should be con- sidered. If the measured field strength in the location in which the Pentacam® / Pentacam® HR is used exceeds the ap- plicable RF compliance level above, the Pentacam® / Pentacam® HR should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Pentacam®...
Page 66
The Pentacam® / Pentacam® HR in intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Pentacam® / Pentacam® HR can help pre- vent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Pentacam®...
(3) The exact obtainable load regulation depends upon the output cord selected and load current. (4) Due to requests in market and advances in technology, specifications subject to change without notice. For the details of safety approval, please consult the factory. 68 / 72 Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00)
License handling from local license server to PEMS and vice versa Storage and data export to local network storage and loading from local network storage Printout to local printer 70 / 72 Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00)
Page 71
Changes to the IT-Network include: changes in IT-Network configuration connection of additional items to the IT-Network disconnecting items from the IT-Network update of equipment connected to the IT-Network Instruction Manual Pentacam® / Pentacam® HR (10028075 - Rev00) 71 / 72...
Page 72
WW W. OCUL US .D E OCULUS Optikgeräte GmbH Postfach • 35549 Wetzlar • GERMANY Phone +49 641 2005-0 • Fax +49 641 2005-255 E-mail: sales@oculus.de • www.oculus.de...
Need help?