Diagnostics and troubleshooting
48
12.8

Event logbook

12.8.1

Event history

To call up the measures to rectify a diagnostic event:
• Via "FieldCare" operating tool →  43
• Via "DeviceCare" operating tool →  43
12.8.2

Overview of information events

Unlike a diagnostic event, an information event is displayed in the event logbook only and
not in the diagnostic list.
Info number
I1000
--------(Device ok)
I1089
Power on
I1090
Configuration reset
I1091
Configuration changed
I1111
Density adjust failure
I1151
History reset
I1157
Memory error event list
I1209
Density adjustment ok
I1221
Zero point adjust failure
I1222
Zero point adjustment ok
I1335
Firmware changed
I1397
Fieldbus: access status changed
I1398
CDI: access status changed
I1512
Download started
I1513
Download finished
I1514
Upload started
I1515
Upload finished
I1622
Calibration changed
I1624
All totalizers reset
I1629
CDI: login successful
I1635
Reset to delivery settings
12.9

Resetting the measuring device

The entire device configuration or some of the configuration can be reset to a defined state
with the Device reset parameter (→  49).
Navigation
"System" menu → Device management → Device reset
Dosimass Modbus RS485