Dosimass Modbus RS485
Flow limit
Pressure loss
Heating
Vibrations
Endress+Hauser
Select the nominal diameter by optimizing between the required flow range and
permissible pressure loss.
For an overview of the full scale values for the measuring range, see the "Measuring
range" section → 56
• The minimum recommended full scale value is approx. 1/20 of the maximum full scale
value
• In most applications, 20 to 50 % of the maximum full scale value can be considered ideal
• A low full scale value must be selected for abrasive media (such as liquids with entrained
solids): flow velocity < 1 m/s (< 3 ft/s).
To calculate the flow limit, use the Applicator sizing tool → 55
To calculate the pressure loss, use the Applicator sizing tool → 55
→ 20
→ 21
Technical data
65