Diagnostics and troubleshooting
46
Diagnostic
Short text
number
492
Frequency output 1 to n
simulation active
493
Pulse output 1 to n
simulation active
494
Switch output 1 to n
simulation active
495
Diagnostic event
simulation active
496
Status input 1 simulation
active
Diagnostic of process
834
Process temperature too
high
835
Process temperature too
low
842
Process value below limit
862
Partly filled pipe
880
Output overloaded
910
Tubes not oscillating
912
Medium inhomogeneous
913
Medium unsuitable
948
Oscillation damping too
high
991
Batch process aborted
992
Batch start failed
1)
Diagnostic behavior can be changed.
12.6
Pending diagnostic events
The Diagnostics menu allows the user to view the current diagnostic event and the
previous diagnostic event separately.
To call up the measures to rectify a diagnostic event:
• Via "FieldCare" operating tool → 43
• Via "DeviceCare" operating tool → 43
Remedy instructions
Deactivate simulation frequency
output
Deactivate simulation pulse output C
Deactivate simulation switch
output
Deactivate simulation
Deactivate status input simulation
Reduce process temperature
Increase process temperature
Low flow cut off active!
Check low flow cut off
configuration
1. Check for gas in process
2. Adjust detection limits
Reduce load at the outputs
1. Check electronic module
2. Check sensor
1. Check process cond.
2. Increase system pressure
1. Check process conditions
2. Check electronic modules or
sensor
Check process conditions
Check process conditions
1. Check fill quantity
2. Check device status
3. Complete last batch
4. Check switch output
configuration
Dosimass Modbus RS485
Status
Diagnostic
signal
behavior
[from the
[from the
factory]
factory]
C
Warning
Warning
C
Warning
C
Warning
C
Warning
1)
S
Warning
1)
S
Warning
1)
S
Warning
1)
S
Warning
S
Warning
F
Alarm
1)
S
Warning
1)
S
Warning
1)
S
Warning
1)
F
Alarm
1)
F
Warning
Endress+Hauser
Need help?
Do you have a question about the Dosimass Modbus RS485 and is the answer not in the manual?