The information contained in this document is subject to change without notice. This document contains proprietary information that is protected by copyright.All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of the Keysight Technologies. Keysight Technologies Japan...
Page 5
Because of the danger of introducing additional hazards, do not install substitute parts or perform unauthorized modifications to the instrument. Return the instrument to a Keysight Technologies Sales and Service Office for service and repair to ensure that safety features are maintained.
Instruments are warranted for a period of one year. Fixtures and adapters are warranted for a period of 90 days. During the warranty period, Keysight Technologies Company will, at its option, either repair or replace products that prove to be defective.
Assistance Product maintenance agreements and other customer assistance agreements are available for Keysight Technologies products. For any assistance, contact your nearest Keysight Technologies Sales and Service Office. Addresses are provided at the back of this manual. Safety Symbol General definitions of safety symbols used on the instrument or in manuals are listed below.
Operation This operating note provides complete information on the 16048E Test Leads. The 16048E is shown in Figure 1-1, its physical dimensions are given in Table 1-1. To order additional copies of this operating note, use the part number listed on the rear...
The 16048E consists of a direct attachment, 4-terminal pair type fixture which is equipped with four BNC (m) terminated-coaxial test leads. These test leads are used to attach user-fabricated test fixtures. Cable length is 4 meter. The 16048E is shown in Figure 1-1.
Operation Specifications Specifications Table 1-1 Specifications of the 16048E 4-terminal pair type fixture which is equipped with four Function: BNC (m) terminated-coaxial test leads. Connector Type: BNC male Cable Length: Approximately 4 m Test Leads Length: Approximately 3.8 m Maximum Voltage: ±...
Compensation for Fixture Residual Impedance Error Compensation for Fixture Residual Impedance Error The 16048E has inherent stray capacitance, residual inductance, and residual resistance that affect the accuracy of measured values. To compensate for, or negate, these residuals to minimize measurement error, the instrument’s Open/Short compensation procedure should be performed.
A BNC bracket (Keysight PN 16032-60001) is furnished with the 16048E. Figure shows an example for using the BNC bracket with the 16048E. By connecting the 16048E to the BNC bracket, the 16048E’s outer shield conductors are connected together to construct the four-terminal pair measurement circuit configuration.
Service Maintenance Maintenance An exploded view of the 16048E (for parts identification) is shown in Figure 2-1 Figure 2-2. Do not disassemble any further than shown. Maintenance consists principally of cleaning contacts and replacing worn or damaged parts. Take special care when cleaning contacts.
Page 19
Service Maintenance Table 2-1 Parts Identification 1/2 Keysight Part Reference Qty. Description 16048-04013 COVER TOP 2360-0192 SCREW 16048-04014 COVER BOTTOM 1250-0089 CONTACT 1250-0052 CONNECTOR BNC 16047-40000 STOPPER 2190-0260 WASHER 2200-0013 SCREW 16048-61612 CABLE ASSEMBLY 0400-0203 GROMET 16048-04015 COVER TOP 2360-0192...
Page 20
Service Maintenance Keysight Part Reference Qty. Description 16032-10022 PLATE 1250-0188 CONNECTOR BNC 2190-0016 WASHER 2950-0001 16032-60001 BNC BRACKET (9 through 12) Chapter 2...
Need help?
Do you have a question about the 16048E and is the answer not in the manual?
Questions and answers