Page 1
Keysight 1000A Series Oscilloscopes Service Guide...
Page 2
Third Edition, April 2013 document or of any information contained to the product or loss of important herein. Should Keysight and the user have a Printed in Available in electronic format only separate written agreement with warranty data. Do not proceed beyond a Keysight Technologies, Inc.
Keysight 1000A Series Oscilloscopes—At a Glance The Keysight 1000A Series oscilloscopes are low-cost portable digital storage oscilloscopes (DSOs) that deliver these powerful features: • Two and four-channel, 60 MHz, 100 MHz, and 200 MHz bandwidth models. • Bright 5.7 inch QVGA (320 x 240) TFT color LCD display and small footprint (to save bench space).
In This Book This guide contains service information for the Keysight 1000A Series oscilloscopes. Service Describes oscilloscope maintanance, performance testing, and what to do if your oscilloscope requires service. Keysight 1000A Series Oscilloscopes Service Guide...
Contents Keysight 1000A Series Oscilloscopes—At a Glance In This Book Figures Tables Service Cleaning the Oscilloscope Testing Performance Before Testing Performance To test DC gain accuracy To test analog bandwidth, maximum frequency To test time scale accuracy To test trigger sensitivity...
Page 6
Contents Keysight 1000A Series Oscilloscopes Service Guide...
Figure 6. Channel 1 Horizontal Scale Setting Figure 7. Averages Menu Item Figure 8. Signal Generator Waveform Figure 9. Connecting Equipment for Time Scale Accuracy Test Figure 10. Connecting Equipment for Trigger Sensitivity Test Keysight 1000A Series Oscilloscopes Service Guide...
Page 8
Figures Keysight 1000A Series Oscilloscopes Service Guide...
Tables Table 1. Keysight 1000A Series Oscilloscope Models Table 2. Equipment Required to Perform All Tests Table 3. DC Gain Accuracy Specification Table 4. Equipment Required for DC Gain Accuracy Test Table 5. Analog Bandwidth Specification Table 6. Equipment Required for Analog Bandwidth Test Table 7.
Page 10
Tables Keysight 1000A Series Oscilloscopes Service Guide...
Keysight 1000A Series Oscilloscopes Service Guide 1 Service Cleaning the Oscilloscope Testing Performance Performance Test Record Returning the Oscilloscope to Keysight for Service Contacting Keysight This chapter describes oscilloscope maintanance, performance testing, and what to do if your oscilloscope requires service.
Do not use too much liquid in cleaning the oscilloscope. Water can enter the CAUTION oscilloscope’s front panel, damaging sensitive electronic components. 3 Make sure that the instrument is completely dry before reconnecting it to a power source. Keysight 1000A Series Oscilloscopes Service Guide...
The procedures are written to minimize the number and types of oscilloscopes and accessories required. The oscilloscopes in these lists are ones that are currently available for sale by Keysight at the time of writing this document. In some cases, the test procedures use features specific to the oscilloscopes in the recommended equipment list.
Type-N (m) to SMA (f) Keysight 1250-1250 Adapter SMA to BNC Adapter Analog Bandwidth SMA (m) to BNC (m) Keysight 1250-0831 Type-N to BNC Adapter Trigger Sensitivity Type-N (m) to BNC (f) Keysight 1250-0780 Keysight 1000A Series Oscilloscopes Service Guide...
Page 15
30 minutes prior to the start of any performance test. Calibration 1 Push the [Utility] key on the front panel. 2 In the Utility menu, press Sel f-Cal. 3 Follow the on-screen instructions. Keysight 1000A Series Oscilloscopes Service Guide...
2 Press the [Defaul t Setup] front panel key. 3 Press the [Acquire] front panel key. 4 In the Acquire menu, press the Acquisition softkey until “Average” appears. 5 Press the Averages softkey and turn the entry knob until “256” appears. Keysight 1000A Series Oscilloscopes Service Guide...
6 Set the channel 1 probe attenuation to 1X. 7 Set the channel 1 vertical sensitivity value to 2 mV/div. 8 Set the power supply to +6 mV. 9 Connect the equipment as shown in Figure Keysight 1000A Series Oscilloscopes Service Guide...
Figure 2 Connecting Equipment for DC Gain Accuracy Test 10 Press [Measure]. 11 Press the Vol tage softkey. 12 Turn the entry knob to select the Vavg measurement as shown below. Keysight 1000A Series Oscilloscopes Service Guide...
DMM- 14 Repeat step 15 through step 23 for channel 2 on two-channel oscilloscope models or for channels 2, 3, and 4 on four-channel models. 15 Set the power supply voltage to +6 mV. Keysight 1000A Series Oscilloscopes Service Guide...
Page 20
Calculate the DC Gain using the following expression and record this value in the DC Gain Test section of the Performance Test Record: ΔV ‚Äì V scope+ scope- -------------- - = ------------------------------------------------ - DCGain = ΔV ‚Äì V DMM+ DMM- Keysight 1000A Series Oscilloscopes Service Guide...
Service To test analog bandwidth, maximum frequency Ensure that the input voltage to the oscilloscope never exceeds 300 Vrms. CAUTION This procedure is the only acceptable method for testing the bandwidth of a 1000A Series NOTE oscilloscope. Table 5 Analog Bandwidth Specification...
50Ω BNC (f) to BNC (m) feed through Keysight 0960-0301 Through Adapter terminator Type-N to SMA Type-N (m) to SMA (f) Keysight 1250-1250 Adapter SMA to BNC Adapter SMA (m) to BNC (m) Keysight 1250-0831 Keysight 1000A Series Oscilloscopes Service Guide...
4 Press the [Defaul t Setup] front panel key. 5 Press the [Auto-Scale] front panel key. 6 Set the channel 1 probe attenuation to 1X. 7 Set the channel 1 vertical scale to 200 mV/div. Keysight 1000A Series Oscilloscopes Service Guide...
Service Figure 5 Channel 1 Vertical Scale Setting 8 Set the horizontal scale to 500 ns/div. Figure 6 Channel 1 Horizontal Scale Setting Keysight 1000A Series Oscilloscopes Service Guide...
Vpp menu item. 15 Set the signal generator to a 1 MHz sine wave with a peak-to-peak amplitude of about 6 divisions as it appears on the oscilloscope screen. Keysight 1000A Series Oscilloscopes Service Guide...
17 Using the power meter reading, convert this measurement to Volts RMS using the expression and record it in the “Performance Test Record" on page 35: × 50Ω meas 1MHz For example, if Pmeas = 3.65 mW: × 3.65 mW 50Ω 427 mV 1MHz Keysight 1000A Series Oscilloscopes Service Guide...
22 Using the power meter reading, convert this measurement to Volts RMS using the expression and record it in the “Performance Test Record" on page 35: × 50Ω meas For example, if Pmeas = 3.65 mW: × 3.65 mW 50Ω 427 mV Keysight 1000A Series Oscilloscopes Service Guide...
Page 28
Record" on page 35. To pass this test, this value must be greater than -3.0 dB. 24 For the remaining channels, move the power splitter to the channel and repeat steps 4 through 23. Keysight 1000A Series Oscilloscopes Service Guide...
+ 5 ppm * (years since manufacture) You can get the “years since manufacture” from the serial number, which is in this format: CN XX XX XXXX Week in Year Year Code (year - 1960) Country Code Keysight 1000A Series Oscilloscopes Service Guide...
2 Connect the output of the signal generator to oscilloscope channel 1 using the BNC cable. Oscilloscope Signal Generator 50W feed through BNC cable Type-N (m) to BNC (f) adapter Figure 9 Connecting Equipment for Time Scale Accuracy Test Keysight 1000A Series Oscilloscopes Service Guide...
Page 31
The number of nanoseconds is equivalent to the time scale error in ppm. f Record the result and compare it to the limits in the “Performance Test Record" on page 35. Keysight 1000A Series Oscilloscopes Service Guide...
Set the input channel’s vertical scale to 5 mV/div. g Decrease the amplitude from the signal generator to the specified vertical division (1.5 div at maximum bandwith or 1 div at 10 MHz). Keysight 1000A Series Oscilloscopes Service Guide...
Page 34
If adjusting the trigger level makes the trigger stable, the test still passes. h Record the result as Pass or Fail in the “Performance Test Record" page 35. Repeat these steps for the remaining oscilloscope channels. Keysight 1000A Series Oscilloscopes Service Guide...
200 mV/div ±600 mV +0.97 to +1.03 500 mV/div ±1.5 V +0.97 to +1.03 1 V/div ±3.0 V +0.97 to +1.03 2 V/div ±6.0 V +0.97 to +1.03 5 V/div ±15.0 V +0.97 to +1.03 Keysight 1000A Series Oscilloscopes Service Guide...
Page 36
200 mV/div ±600 mV +0.97 to +1.03 500 mV/div ±1.5 V +0.97 to +1.03 1 V/div ±3.0 V +0.97 to +1.03 2 V/div ±6.0 V +0.97 to +1.03 5 V/div ±15.0 V +0.97 to +1.03 Keysight 1000A Series Oscilloscopes Service Guide...
Test Limits ≥ 5 mV/div: 1.5 division at maximum bandwidth ≥ 5 mV/div: 1 division at 10 MHz < 5 mV/div: 1.5 division at 20 MHz < 5 mV/div: 1 division at 10 MHz Keysight 1000A Series Oscilloscopes Service Guide...
4 Pack the oscilloscope in foam or other shock absorbing material and place it in a strong shipping container. You can use the original shipping materials or order materials from an Keysight Technologies Sales Office. If neither are available, place 8 to 10 cm (3 to 4 inches) of shock-absorbing material around the oscilloscope and place it in a box that does not allow movement during shipping.
Service Contacting Keysight Information on contacting Keysight Technologies can be found at www.keysight.com/find/contactus. Keysight 1000A Series Oscilloscopes Service Guide...
Keysight 1000A Series Oscilloscopes Service Guide A Safety Notices Warnings Safety Symbols This apparatus has been designed and tested in accordance with IEC Publication 1010, Safety Requirements for Measuring Apparatus, and has been supplied in a safe condition. This is a Safety Class I instrument (provided with terminal for protective earthing).
Hazardous voltage symbol. Earth terminal symbol: Used to indicate a circuit common connected to grounded chassis. Keysight 1000A Series Oscilloscopes Service Guide...