Test fixture

Advertisement

Quick Links

Keysight 16092A Test Fixture
Operation and
Service Manual

Advertisement

loading
Need help?

Need help?

Do you have a question about the 16092A and is the answer not in the manual?

Questions and answers

Summary of Contents for Keysight 16092A

  • Page 1 Keysight 16092A Test Fixture Operation and Service Manual...
  • Page 2 EULA. governed by United States and Declaration of Conformity Keysight shall be under no obligation international copyright laws. to update, revise or otherwise modify Trademark Acknowledgments Declarations of Conformity for this the Software.
  • Page 3: Table Of Contents

    Maintenance ..............7 Keysight 16092A Test Fixture Operation and Service Manual...
  • Page 4 Contents Keysight 16092A Test Fixture Operation and Service Manual...
  • Page 5: Operation

    Operation and Service Manual Operation This operating note provides the information required to maintain and repair the 16092A Spring Clip Fixture. The 16092A pictorially shown in Figure 1-1 was specially designed for use with the Impedance analyzer. How to use the 16092A, performance characteristic data, and operating instructions are given in the Impedance Analyzer's operation and service Manual.
  • Page 6: Description

    Under this condition, the residual inductance and resistance are displayed. The left and right spring clip assemblies on the 16092A spring Clip Fixture are composed of the same parts and are assembled and disassembled in the same way as shown in the figure.
  • Page 7: Service

    Maintenance Figure 2-1 lists all replaceable parts of the 16092A. To identify the locations of the individual parts listed refer to the exploded view. Do not disassemble any further than shown in the figure. To order the necessary replacement parts, identify them by their Keysight Technologies part number.
  • Page 8 Service Maintenance Figure 2-1 Parts Identification Keysight 16092A Test Fixture...
  • Page 9 0520-0133 SCREW SCREW Set Change: 0515-0661 2190-0103 WASHER WASHER Set change: 3050-1066 16092-40010 HOLDER 16092-40010 HOLDER 16092-24012 SCREW 16092-24012 SCREW 1460-0352 SPRING 16092-29001 SPRING (16092-29001) 16092-40011 LEVER 16092-01201 LEVER (16092-01201) 16092-40016 HOLDER HOLDER Set Change: 16092-40616 Keysight 16092A Test Fixture...
  • Page 10 SCREW 16092-24013 SCREW 16092-24012 SCREW 16092-24012 SCREW 16092-08010 GROUND 16092-08010 GROUND SPRING SPRING 1540-0622 CASE 9300-2603 CASE 16092-40018 CENTER POST 16092-40019 CENTER POST (16092-40019) WITH WITH INSULATOR INSULATOR 16092-24014 GROUND GROUND Set change: SCREW SCREW 16092-24614 Keysight 16092A Test Fixture...
  • Page 11 REPLACEMENT OF THE APC-7 CONNECTOR CONTACT. To maintain optimum contact between the APC-7 connector contacts it is recommended that the RF Connector Contact (Keysight Part No. 1250-0907) of the center conductor be replaced every 50 connections/disconnections or as required (when damaged or worn). A Pin vise of the 11591A APC-7 Pin Replacement Kit is required to remove the contact from the connector.
  • Page 12 Do not use aromatic or chlorinated hydrocarbons, esters, ethers, terpenes, higher alcohols, ketones, or such chemicals as benzene, toluene, turpentine, dioxane, gasoline, cellulose acetate, or carbontetrachloride. Keep exposure of the connector parts to both the cleaning fluid and its vapors as brief as possible. Keysight 16092A Test Fixture...
  • Page 13 This information is subject to change without notice. © Keysight Technologies 1986-2017 Edition 5, October 2017 *16092-90010* 16092-90010 www.keysight.com...