Keysight 11683AZ Option K01 and K02 Power Meter Range Calibrator Installation Note Special Supplement Description Assembly Process Operation Performance Test...
Page 10
Special Supplement Description The 11683AZ-K01/11683AZ-K02 is a Field Upgrade of a standard 11683A range calibrator to 11683AZ-H01. This kit is used to modify the unit so that it operates with either the Internal DC-Reference source, or an external programmable DC-Reference source.
Page 11
50 mm Blue Wire 2950-0043 2190-0016 Washer 11683-00011 Ground Connector 8150-0451 130 mm Yellow Wire 8150-0454 200 mm Violet Wire 11683-00005 Xmer Mounting Bracket 11683-60004 Power Switch Assembly 2360-0115 Screw 2420-0001 Hexagonal Nut 11683-20008 Insulator Keysight 11683AZ Option K01 and K02 Installation Note...
Page 12
6 Remove resistor (196K?) that is between yellow and violet wire (8150-0454) from multi-position switch (11683-60005) 7 Keep violet wire (8150-0454) on multi-position switch (11683-60005) but remove yellow wire (8150-0451) from multi-position switch (11683-60005) for next step. Keysight 11683AZ Option K01 and K02 Installation Note...
Page 13
20 Solder black wire (8150-0447) from bottom BNC lug (0360-1190) to 31.62K resistor on multi-position switch (11683-60005). 21 Ensure yellow wire from middle of SPDT switch (3101-0163) to the feed through going to the power sensor housing was solder. Keysight 11683AZ Option K01 and K02 Installation Note...
Page 14
The above calculated voltages for range 100 mW to 300 W are based on the table shown in the 11683A Operating and Service Manual (11683-90014). The remaining voltages are based in the Keysight 11683AZ Option K01 and K02 Installation Note...
Page 15
4 Connect the jumper between the DC REF INPUT BNC wire and the RANGE switch wire using an alligator test clip. 5 Set the REF SELECT switch to EXT. 6 Connect the 11683AZ-H01 Range Calibrator to the power line and turn it ON. Keysight 11683AZ Option K01 and K02 Installation Note...
Page 16
10 Disconnect the 4-wire cable and jumper from the 11683A. This concludes the external DC-reference circuit performance test. If any of the voltage or resistance readings are incorrect, refer to the troubleshooting information in the 11683A Operating and Service Manual (11683-90014). Keysight 11683AZ Option K01 and K02 Installation Note...
Page 17
11683-20008 Insulator In other words, the 11683AZ Option H01 is similar to the standard 11683A and the information in the Operating and Service Manual for the standard 11683A applies to this instrument. Keysight 11683AZ Option K01 and K02 Installation Note...
Page 18
Special Supplement Figure 1-2 Range Switch/Sampling Gate Schematic Diagram Keysight 11683AZ Option K01 and K02 Installation Note...