OPERATION AND
MAINTENANCE MANUAL

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AEU-350 and is the answer not in the manual?

Questions and answers

Summary of Contents for Aseptico AEU-350

  • Page 1 AEU-350 / AEU-350S Ultra-Portable Dental System (AHP-72MK-FO)* (Scaler included with AEU-350S)* (AHP-101)* (AHP-53)* *Shown with optional equipment OPERATION AND MAINTENANCE MANUAL...
  • Page 2 AEU-350 / AEU-350S ! SAFETY PRECAUTIONS Aseptico accepts no liability for direct or consequential injury or dam- s CAUTION: Do not allow oil from the handpiece to drain into the age resulting from improper use, arising in particular through the motor.
  • Page 3: Electromagnetic Compatibility

    AEU-350 / AEU-350S ELECTROMAGNETIC COMPATIBILITY This equipment meets all requirements for safety and performance, related to Electromagnetic Compatibility Standard IEC 60601-1-2:2014. s NOTE: The emissions characteristics of this equipment make it suit- able for use in industrial areas and hospitals (CISPR 11 class A). If it is...
  • Page 4: Table Of Contents

    AEU-350 / AEU-350S Congratulations! TABLE OF CONTENTS ELECTROMAGNETIC COMPATIBILITY Your new Aseptico® GO AEU-350 / AEU-350S system is the finest ultra-portable electric dental system available to the dental profes- TABLE OF CONTENTS sion. The AEU-350S includes an ultrasonic scaler with irrigation, and INTENDED PURPOSE &...
  • Page 5: Intended Purpose & Intended Users

    (AEU-350S only). Intended Users: The AEU-350 and AEU-350S are prescription devices that are used by healthcare professionals in the field of dentistry. The optional scaler should only be used by licensed dental profes- sionals trained in the proper use of scaling devices.
  • Page 6: Package Contents

    AEU-350 / AEU-350S PACKAGE CONTENTS q AEU-350/S Unit (P/N 120426 or 120443) q Footswitch (P/N AE-7PM) q HVE (High-Volume Ejector) Valve and Hose Assembly (P/N AA-350-HVE) q Power Cord (P/N 840107) q Operation and Maintenance Manual q Waste Collection Tank (P/N 330811) (P/N 421222)
  • Page 7: Sold Separately

    AEU-350 / AEU-350S PACKAGE CONTENTS SOLD SEPARATELY q Vacuum Line Filter (x2) (P/N AA-350-VLF) q Scaler Handpiece Assembly (P/N 710658) q 24 VDC Battery Power Cord (AEU-350S only, replaceable) (Optional equipment) q Scaler Cable Assembly (P/N 710659) (AEU-350S only, permanently attached)  View the DC input socket on your unit to determine which 24 VDC battery power cord to purchase.
  • Page 8 AEU-350 / AEU-350S  The EXPANDED SETUP has the advantage of making the instruments easy to reach, even from a standing position. It has the disadvantage of not having automatic vacuum actuators in the HVE and saliva ejector secondary instrument holders, requiring you to toggle the Vacuum Enable switch when suction is needed.
  • Page 9 AEU-350 / AEU-350S  The COMPACT SETUP has the advantages of a quicker setup and automatically actuated HVE and saliva ejector instrument holders. Electric Motor (High Volume Ejector) (Handpiece not included) Saliva Ejector (Low Volume Ejector) Air/Water Syringe Scaler Electric Motor / Scaler Settings...
  • Page 10: System Setup

    Note that the syringe lines  are permanently con- nected to the bulkhead. The AEU-350 features an auxiliary water port  , and the AEU-350S features a scaler which is per- manently attached at this location  ...
  • Page 11 AEU-350 / AEU-350S Plug the electric motor cord  into its socket in the bulkhead. Lift up the unit to the vertical position with the door open. Place the footswitch  where it is convenient to operate.  The plug can only be inserted into the socket in one orientation.
  • Page 12 AEU-350 / AEU-350S SYSTEM SETUP (continued) 11 Fill the water bottle  with water (never saline). Insert the uptake tube  into the bottle, then screw the bottle  into the fitting on the bulkhead. AEU-350S only: Place the front lip  of the scaler holder on the edge of the door then rotate until the back lip ...
  • Page 13 AEU-350 / AEU-350S 13 Insert the syringe tip  into the syringe, a 1:1 or 1:5 handpiece 15 Optional 24VDC Operation: Unplug the AC power cord  onto the micromotor, and disposable tips onto the saliva and remove the DC lockout cover ...
  • Page 14: Control Panel

    Touch Screen - Scaler Mode (AEU-350S Only)  When your AEU-350 unit is shut off and turned back on, it will return to the mode and settings from last use. Follow Customize Preset Settings for directions to revert to preset settings.
  • Page 15 Rotate counter-clockwise to increase flow and clockwise to  When your AEU-350 unit is shut off and turned back on, it decrease flow. will return to the mode and settings from last use. Follow Customize Preset Settings for directions to revert to WARNING: Do not use the ultrasonic scaler dry.
  • Page 16: General Operation

    AEU-350 / AEU-350S GENERAL OPERATION Using HVE and Saliva Ejector Vacuums Use the vacuum enable switch to automatically activate the After the unit has been set up and you have made yourself famil- vacuum when the handpieces are not in the holder.
  • Page 17: Maintenance And Sterilization

    AEU-350 / AEU-350S When switched on, the scaler light will illuminate when the MAINTENANCE AND STERILIZATION footswitch is pressed. Failure to adhere to the following instructions or blatant misuse of To use the scaler, the device and/or related components will be considered abuse of the device and will void the warranty.
  • Page 18 AEU-350 / AEU-350S MAINTENANCE AND STERILIZATION O-Rings (cont) Periodically apply PTFE lubricant to all o-rings to keep the system Air/Water Syringe and Scaler Tips operating at peak efficiency. Damaged or worn o-rings can leak and degrade performance; replace as needed. Bulkhead o-rings The syringe features quick-change autoclavable tips.
  • Page 19 AEU-350 / AEU-350S CAUTION: Do not use ultrasonic cleaners. Sterilize Select one of the following sterilization methods: CAUTION: Do not use cold sterilization or dry heat. Gravity Packaged: Sterilize the motor and cord assembly using any one of the steriliza- •...
  • Page 20 Filter, but it is recommended to keep in place to prevent waste aerosols from entering into the lower assembly.  Call Aseptico at 1-800-426-5913 with any questions on sterilization procedures. The water supply bottle should be cleaned and dried after every daily use, and before storage.
  • Page 21: Repacking

    AEU-350 / AEU-350S REPACKING TROUBLESHOOTING Remove all water/waste and clean/sterilize the unit as Unit does not power up: instructed in Maintenance and Sterilization. • Check power cord connection. • Check that AC-DC input selection switch is in correct position. Unplug the power cord and coil.
  • Page 22: Warranty

    Insufficient scaler fluid (AEU-350S only): in part or whole. Aseptico shall be the sole arbiter of such action. • Open the water adjustment knob completely. In the event of alleged defect under warranty, the purchaser •...
  • Page 23: Symbol Descriptions

    AEU-350 / AEU-350S SYMBOL DESCRIPTIONS Scaler Coolant Consult Instructions For Use Water Caution Waste Type B Applied Part Model Number  Important information UDI - Unique Device Identifier Alternating Current Country of Manufacture - USA Direct Current Rx Dental Use Only...
  • Page 24: Replacement Parts

    REPLACEMENT PARTS SPECIFICATIONS Call Aseptico at 1-800-426-5913 to order replacement parts. CASE DIMENSIONS: 34.6 cm W x 23.6 cm D x 54.9 cm H (13.6” x 9.3” x 21.6”) UNIT WEIGHT: 32 lb (14.5 kg), 34 lb (15.5 kg) for AEU-350S...
  • Page 25: Messages

    AEU-350 / AEU-350S MESSAGES NOTES System Messages When the system is powered on, the software part number and version will be displayed briefly on the touch screen. The waste tank sensor will flash the version number of soft- ware in the lower chamber. One flash indicates version 1, two flashes indicates version 2, etc.
  • Page 26: Notes

    AEU-350 / AEU-350S NOTES...
  • Page 27 AEU-350 / AEU-350S NOTES...
  • Page 28 P.O. Box 1548, Woodinville, WA 98072 8333 216th Street S.E., Woodinville, WA 98072 Phone (425) 487-3157 Web: www.aseptico.com Email: info@aseptico.com P/N 421222 • Rev. L • ECO 15926 • 6/2024 • Printed in the USA...