10.5.3 Configuration of the measuring range "M range"

> Configuration of the measuring range:
First select the "M range">, then the measuring range.
The measuring range has 7 settings on the meter.
Measuring range
1)
200µS
1)
2000µS
1)
20mS
1)
500mS
2) 3)
20µS
2) 4)
2µS
2) 3)
200µS
Auto
5)
TDS
5)
Resist.
1)
Measuring range switching possible
2)
Measuring range switching requires a special sensor (measuring ranges are purest water).
This sensor must be ordered separately.
3)
Used cell constant 0,1 cm
4)
Used cell constant 0,01 cm
5)
Measuring range switching also takes place automatically
Via (CONF +1) or (LOG -1) you selected the desired setting.
(ON/OFF) confirms and saves your setting.
(MEAS) terminates the entry dialog without transferring the value.
Main measured variable
LF
LF
LF
LF
LF
LF
LF
> automatically selects the suitable measuring range
> shows the TDS value as a secondary variable
> shows the resistance value as a secondary variable
-1
-1
Secondary measured variable
Saline
Saline
Saline
Saline
Resist
Resist
Resist