Instructions for use SMARTmatic PROPHY S19 – 1.011.6740, SMARTmatic PROPHY S19 K – 1.011.6741 Table of contents Table of contents 1 User instructions....................... 2 Safety ..........................2.1 Infection hazard......................7 2.2 Improper use ....................... 7 2.3 Technical condition......................7 2.4 Accessories and combination with other equipment ............8 2.5 Qualification of personnel....................
Instructions for use SMARTmatic PROPHY S19 – 1.011.6740, SMARTmatic PROPHY S19 K – 1.011.6741 1 User instructions 1 User instructions Dear user, Congratulations on purchasing this KaVo quality product. By following the notes below you will be able to work smoothly, economically and safely.
Page 5
Instructions for use SMARTmatic PROPHY S19 – 1.011.6740, SMARTmatic PROPHY S19 K – 1.011.6741 1 User instructions Serial number UDI symbol Manufacturer Note: Please note accompanying documents Follow the electronic instructions for use HIBC Code CE mark for medical devices...
Page 6
Instructions for use SMARTmatic PROPHY S19 – 1.011.6740, SMARTmatic PROPHY S19 K – 1.011.6741 1 User instructions WARNING In cases which – if not prevented – can lead to death or severe injury. CAUTION In cases which – if not prevented – can lead to minor or moderate injury.
Instructions for use SMARTmatic PROPHY S19 – 1.011.6740, SMARTmatic PROPHY S19 K – 1.011.6741 2 Safety | 2.1 Infection hazard 2 Safety NOTE All serious events occurring in relation to the product must be reported to the manufacturer and the competent authority of the member state, in which the user and/or patient resides.
Instructions for use SMARTmatic PROPHY S19 – 1.011.6740, SMARTmatic PROPHY S19 K – 1.011.6741 2 Safety | 2.4 Accessories and combination with other equipment ▪ Damage (e.g., from dropping) ▪ Irregular running noise ▪ Excessive vibration ▪ Overheating ▪ Insufficient retention force...
Page 9
Instructions for use SMARTmatic PROPHY S19 – 1.011.6740, SMARTmatic PROPHY S19 K – 1.011.6741 2 Safety | 2.6 Service and repair 4 Have the service and testing tasks carried out according to the Medical Device Operator Ordinance. 4 KaVo recommends specifying in-house service intervals where the medical device is brought to a professional shop for cleaning, servicing and a func- tion check.
Instructions for use SMARTmatic PROPHY S19 – 1.011.6740, SMARTmatic PROPHY S19 K – 1.011.6741 3 Product description | 3.1 Intended use 3 Product description SMARTmatic PROPHY S19 (Mat. no. 1.011.6740) SMARTmatic PROPHY S19 K (Mat. no. 1.011.6741) 3.1 Intended use Indications for use: This medical device is: ▪...
Instructions for use SMARTmatic PROPHY S19 – 1.011.6740, SMARTmatic PROPHY S19 K – 1.011.6741 4 Startup and shut down 4 Startup and shut down WARNING Hazard from contaminated products. Patients, users or third parties could be infected by contaminated medical devices.
Instructions for use SMARTmatic PROPHY S19 – 1.011.6740, SMARTmatic PROPHY S19 K – 1.011.6741 5 Operation | 5.1 Attaching the medical device 5 Operation WARNING Rotating dental bur. Cuts, infection and burn injury. 4 Do not touch the dental tool while it is rotating.
Instructions for use SMARTmatic PROPHY S19 – 1.011.6740, SMARTmatic PROPHY S19 K – 1.011.6741 5 Operation | 5.3 Attach the prophylaxis head 5.3 Attach the prophylaxis head NOTE Only use polishing attachments with a Doriot connection that conform to DIN...
Malfunction or complete failure. 4 Do not use Vaseline or other grease or oil. 4 Spray a lint-free cloth with KaVo Spray and apply it to the O-rings on the coupling. 4 Press the O-ring between your fingers to form a loop.
Page 16
Instructions for use SMARTmatic PROPHY S19 – 1.011.6740, SMARTmatic PROPHY S19 K – 1.011.6741 6 Checking for malfunctions and troubleshooting | 6.2 Troubleshooting 4 Spray a lint-free cloth with KaVo Spray, apply it to the new O-rings and in- sert them in the recesses.
Instructions for use SMARTmatic PROPHY S19 – 1.011.6740, SMARTmatic PROPHY S19 K – 1.011.6741 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.1 Preparations at the site of use 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 7.1 Preparations at the site of use...
Instructions for use SMARTmatic PROPHY S19 – 1.011.6740, SMARTmatic PROPHY S19 K – 1.011.6741 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.2 Manual reprocessing CAUTION Never reprocess this medical device in an ultrasonic cleaner. Malfunction and material damage.
Instructions for use SMARTmatic PROPHY S19 – 1.011.6740, SMARTmatic PROPHY S19 K – 1.011.6741 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.3 Automated reprocessing 4 Allow the disinfectant to act for 5 minutes. 7.2.3 Manual drying Use KaVo DRYspray for subsequent drying of the air, water and gear unit ducts.
Instructions for use SMARTmatic PROPHY S19 – 1.011.6740, SMARTmatic PROPHY S19 K – 1.011.6741 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.4 Care products and systems - Servicing NOTE Please comply with the instructions for use of the washer disinfector.
Instructions for use SMARTmatic PROPHY S19 – 1.011.6740, SMARTmatic PROPHY S19 K – 1.011.6741 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.5 Packaging Servicing the chucking system KaVo recommends servicing the chuck system once weekly using the chuck servicing programme integrated in the device.
Instructions for use SMARTmatic PROPHY S19 – 1.011.6740, SMARTmatic PROPHY S19 K – 1.011.6741 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.7 Storage Sterilisation parameters: Select a suitable process from the following sterilisation processes (depending on the available steriliser): ▪...
Instructions for use SMARTmatic PROPHY S19 – 1.011.6740, SMARTmatic PROPHY S19 K – 1.011.6741 8 Optional aids and consumables 8 Optional aids and consumables Available from specialised dental dealers. Material summary Mat.No. INTRA Instrument stand 3.005.5204 Cleanpac 10 units 0.411.9691 Cellulose pad 100 units 0.411.9862...
Instructions for use SMARTmatic PROPHY S19 – 1.011.6740, SMARTmatic PROPHY S19 K – 1.011.6741 9 Terms and conditions of warranty 9 Terms and conditions of warranty This KaVo medical device is subject to the following warranty conditions: KaVo grants the end customer a warranty of proper function and guarantees...
Need help?
Do you have a question about the SMARTmatic PROPHY S19 K and is the answer not in the manual?
Questions and answers