KaVo SURGmatic S11 L Instructions For Use Manual

KaVo SURGmatic S11 L Instructions For Use Manual

Hide thumbs Also See for SURGmatic S11 L:

Advertisement

Quick Links

Instructions for use
SURGmatic S11 L - 1.009.1010
SURGmatic S11 C - 1.009.1005

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SURGmatic S11 L and is the answer not in the manual?

Questions and answers

Summary of Contents for KaVo SURGmatic S11 L

  • Page 1 Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005...
  • Page 2 Distributed by: Manufacturer: KaVo Dental Technologies, LLC Kaltenbach & Voigt GmbH 11727 Fruehauf Drive Bismarckring 39 Charlotte, NC 28273 USA D-88400 Biberach Phone: 847 550 6800 www.kavo.com Fax: 847 550 6825...
  • Page 3: Table Of Contents

    Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005 Table of contents Table of contents 1 User instructions ........................ 4 2 Safety..........................7 2.1 Infection hazard ......................7 2.2 Technical condition......................7 2.3 Accessories and combination with other equipment ............7 2.4 Qualification of personnel....................
  • Page 4: User Instructions

    Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005 1 User instructions 1 User instructions Dear User Congratulations on purchasing this KaVo quality product. By following the in- structions below you will be able to work smoothly, economically and safely.
  • Page 5 Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005 1 User instructions Thermodisinfectable Information on the packaging Material number Serial number Legal Manufacturer CE mark according to Medical Devices Directive EC 93/42 Please note the electronic instructions for use...
  • Page 6 Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005 1 User instructions CAUTION In cases which – if not prevented – could lead to minor or moderate injury. NOTICE In cases which – if not prevented – could lead to material damage.
  • Page 7: Safety

    Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005 2 Safety | 2.1 Infection hazard 2 Safety The instructions for use are a component of the product and must be read care- fully prior to use and be accessible at all times.
  • Page 8: Qualification Of Personnel

    Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005 2 Safety | 2.4 Qualification of personnel ▶ Only use accessories that have been approved for combination with the product by the manufacturer. ▶ Do not make any modifications to the device unless these have been ap- proved by the manufacturer of the product.
  • Page 9: Product Description

    Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005 3 Product description | 3.1 Purpose – Proper use 3 Product description SURGmatic S 11 L SURGmatic S11 L (Mat. no. 1.009.1010) SURGmatic S 11 C SURGmatic S11 C (Mat. no. 1.009.1005) 3.1 Purpose –...
  • Page 10: Transportation And Storage Conditions

    Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005 3 Product description | 3.3 Transportation and storage conditions The handpiece can be mounted on all INTRAmatic (LUX) motors, and motors with a connection in accordance with EN ISO 3964.
  • Page 11: Start Up And Shut Down

    Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005 4 Start up and shut down | 4.1 Checking the amount of water 4 Start up and shut down WARNING Hazard from non-sterile products. Infection hazard for dentist and patient.
  • Page 12: Operation

    Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005 5 Operation | 5.1 Attach the medical device 5 Operation 5.1 Attach the medical device WARNING Detachment of the medical device during treatment. A medical device that is not properly locked in place can become disconnected from the motor coupling and fall off.
  • Page 13: Remove The Handpiece Or Contra-Angle Drill Bit

    Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005 5 Operation | 5.4 Remove the handpiece or contra-angle drill bit WARNING Use of unauthorised cutters or grinders. Injury to the patient or damage to the medical device.
  • Page 14: Troubleshooting

    Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005 6 Troubleshooting | 6.1 Check for malfunctions 6 Troubleshooting 6.1 Check for malfunctions CAUTION Heating of the product. Burns or product damage from overheating. ▶ Do not use the product if it is irregularly heated.
  • Page 15: Reprocessing Steps In Accordance With En Iso 17664

    Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005 7 Reprocessing steps in accordance with EN ISO 17664 | 7.1 Preparation at the site of use 7 Reprocessing steps in accordance with EN ISO 17664 7.1 Preparation at the site of use WARNING Hazard from contaminated products.
  • Page 16: Manual Reprocessing

    Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005 7 Reprocessing steps in accordance with EN ISO 17664 | 7.3 Manual reprocessing 7.3 Manual reprocessing WARNING Sharp tool in the medical device. Injury hazard from sharp and/or pointed tool.
  • Page 17: Automated Internal And External Cleaning And Internal And External Disinfection

    Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005 7 Reprocessing steps in accordance with EN ISO 17664 | 7.5 Care products and systems - Servicing 7.4.1 Automated internal and external cleaning and internal and external disinfection...
  • Page 18: Servicing With Kavo Quattrocare Plus

    Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005 7 Reprocessing steps in accordance with EN ISO 17664 | 7.5 Care products and systems - Servicing ▶ Plug the product onto the cannula, and press the spray button for one second.
  • Page 19: Servicing With Kavo Quattrocare

    Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005 7 Reprocessing steps in accordance with EN ISO 17664 | 7.6 Packaging Note Close the chuck service mode. Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A, close the front door and start theservice procedure.
  • Page 20: Sterilisation

    Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005 7 Reprocessing steps in accordance with EN ISO 17664 | 7.7 Sterilisation 7.7 Sterilisation Sterilisation in a steam steriliser (autoclave) in accordance with ISO 17665-1 CAUTION Premature wear and malfunctions from improper servicing and care.
  • Page 21: Tools

    Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005 8 Tools 8 Tools Available from dental suppliers. Material summary Mat.No. Instrument stand 3.005.5204 Cleanpac 10 units 0.411.9691 Cellulose pad 100 units 0.411.9862 Bur stop 0.524.0892 Hook 0.410.1963...
  • Page 22: Terms And Conditions Of Warranty

    Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005 9 Terms and conditions of warranty 9 Terms and conditions of warranty The following warranty conditions apply to this KaVo medical device: KaVo provides the end customer with a warranty of proper function and guar-...

This manual is also suitable for:

Surgmatic s11 c1.009.10101.009.1005

Table of Contents