Acelity V.A.C.ULTA Maintenance Manual page 13

Therapy system

Advertisement

Export Logs Screen
V.A.C.® Therapy
Mode:
Continuous
QC Checklist
port
Logs
NNT
]
Upgrade Software
|
Application Version: 0
|
Language Version: 0
> Operating
System Version0
|
Status
| Complete
V.A.C.® Therapy
Mode: Continuous
QC Checklist
REN]
Export Logs
NINE
]
=
Upgrade Software
ビバ
Application Version: 0
Language Version: 0
7 Operating System Version 0
©
E- Preventative Maintenance
(FU
Restore Patient Defaults
| complete
|
ST
1.
Open the User Data Interface door at the front of
the V.A.C.ULTA™ Therapy Unit.
2.
Insert memory device into the unit data port.
3.
Press the Arrow button displayed next to
Export Logs.
4.
When export is complete, remove memory device
from data port and close the User Data Interface
door.
If a red X appears during export, remove the memory
device, reinsert into the data port and press the Arrow
button again. If a red X appears after a second attempt,
refer to the Data Transfer Errors section in the
V.A.C.ULTA™ Therapy Unit User Manual. If errors
continue, contact KCI (see page 30).