V.A,C. VERAFLO" Therapy
a
ili
iii
AE
_
Go
à Instill Pump Performance
Pump Duty: 0
|
J Motor speed: 0
00:00:00
'VAC. VERALINK™ Cassette NotEngaged |
Alarm
/
Solution Bag / Bottle Empiy Alarm
(i) VAC. VERAFLO" Blockage Alarm |
00:00:00
Back
ac 'Checklist
Mode: Continuous
QC Checklist
Logs
Version: 0
Language Version: 0
Version: 0
| Complete
5.
Ifthe value continues to display > 88%
。
Press and hold the Power button until the display turns off
to turn the V.A.C.ULTA™ Therapy Unit off.
+ .
Verify that all plugs are undamaged and firmly attached to
the V.A.C.ULTA™ Therapy Unit.
。
Press and hold the Power button until the light comes on to
turn the V.A.C.ULTA™ Therapy Unit on.
*
Verify the V.A.C.ULTA" Therapy Unit is receiving power when
plugged into the AC wall outlet.
«
Press the QC Checklist button on the Startup screen.
ㆍ
Press the Advanced QC button from the QC Checklist
screen.
*
Type the Advanced QC Code M8G77R in the Access Code
window on the Znter Access Code screen and press the OK
button.
。
Press the Arrow button displayed next to V.A.C. VERAFLO""
Therapy from the QC Check/ist screen.
。
Press the Arrow button displayed next to Instill Pump
Performance on the A.C. VERAFLO™ Therapy screen.
«If after powering down and re-running test the Pump Duty
value is still greater than 88%, contact KCI (see page 30).
6.
If the Pump Duty value is < 88% press the QC
Checklist button to return to the OC Check/ist
screen.
7.
Press the Arrow button displayed next to Restore
Patient Defaults.
レー ニン ーー ニーー
Complete
8.
Press the Complete button.