Acelity V.A.C.ULTA Maintenance Manual page 15

Therapy system
Hide thumbs Also See for V.A.C.ULTA:

Advertisement

4.
Press the Arrow button displayed next to Pressure
V.A.C.® Therapy
Accu racy.
Mo
ontinuous
V.A.C.© Therapy
а —
Blockage Alarm
D | Canister NotEngaged
Alam = | о
077
>
QC Checklist
"Next
5.
Verify the manometer reading is +/- 10mmHg of 125 mmHg and
record the number on the V.A.C.ULTA™ Therapy Unit Checklist
(see this manual).
V.A.C.® Therapy
Mode: Continuous
6.
The unit has passed the test a green check mark will be
displayed in the Status box. If the unit fails the Pressure
Calibration screen will appear automatically.
7.
Continue to the V.A.C.® Therapy Alarm Testing section
Blockage
Canister Not Engaged Alarm