Qc Checklist
Warning:
Running the QC Checklist will result in patient
history being deleted.
Once deleted, data cannot be recovered.
Enter Access Code Screen
Enter Access Code
Access Code
(Requires minimum of 6 characters)
8.
Press the Advanced QC button on the
OC Checklist screen.
Running the QC Checklist will result in patient
history being deleted. Once deleted, data cannot
be recovered.
9.
Use the keypad to type the Advanced QC Code
M8G77R
into the Access Code window on the
Enter Access Code screen.
Press the OK button.
If the Export Log procedure on the following page
is not performed all information on the unit will
be lost once the defaults are restored at the end of
this checklist. Once deleted, this data cannot be
recovered.