Maintenance And Cleaning; Accessories - IKA C-MAG HS 7 control Instructions Manual

Hide thumbs Also See for C-MAG HS 7 control:
Table of Contents

Advertisement

Maintenance and cleaning

› The device is maintenance-free� It is only subject to the natural wear and tear of components
and their statistical failure rate�
/// Cleaning:
› For cleaning disconnect the mains plug!
› Use only cleaning agents which have been approved by IKA to clean IKA devices�
These are: water (containing surfactant) and isopropyl alcohol�
› Wear protective gloves during cleaning the devices�
› Electrical devices may not be placed in the cleansing agent for the purpose of cleaning�
› Do not allow moisture to get into the device when cleaning�
› Before using another than the recommended method for cleaning or decontamination, the
user must ascertain with IKA that this method does not destroy the device�
/// Spare parts order:
› When ordering spare parts, please give:
- device type�
- serial number, see type plate�
- position number and description of spare part, see www.ika.com�
- software version�
/// Repair:
› Please only send devices in for repair that have been cleaned and are free of materials which
might present health hazards�
› For repair, please request the "Safety Declaration (Decontamination Certificate)" from
IKA or use the download printout of it from IKA website at www.ika.com�
› If your appliance requires repair, return it in its original packaging� Storage packaging is not
sufficient when sending the device - also use appropriate transport packaging�

Accessories

› For accessories see www.ika.com�
32
Error codes
› The fault is shown by an error code on the display as following if the error occurs�
Proceed as follows in such cases:
- Turn off the device by using the main switch (left « on, right « off)�
- Carry out corrective measures�
- Restart the device�
Error code | Causes | Effect | Solutions
Er02 - Watchdog error
Causes
PC does not transmit any data within the set watchdog time
connection to PC interrupted
Effect
heating switched off
motor switched off
Solutions
change watchdog time
transmit data from PC within set watchdog time (OUT_WDx@m)
check cable and plug
Er03 - Temperature inside device is higher than 80 °C
Causes
heat accumulation between heating plate and housing
permitted ambient temperature exceeded
Effect
heating switched off
Solutions
switch device off, leave to cool and switch on again
change experiment
observe maximum permissible ambient temperature
Er04 - Motor control is unavailable
Causes
motor blocked or overloaded
Effect
heating switched off
motor switched off
Solutions
reduce load torque or use smaller magnetic rods
reduce target speed
Er05 - No temperature increase measured by temperature sensor
(selected time in menu)
Causes
sensor not in medium
volume of medium to be measured too large
heat conductivity of medium to be measured too low
heat conductivity of the vessel is too low
in the case of indirect heating, the overall heat conductivity resistance is too
large
Effect
heating switched off
Solutions
place the sensor in the medium
reduce the volume of the media
use a carrier fluid with better heat conductivity properties
replace the glass vessel with a metal pot
increase the "Time-out" period
33

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the C-MAG HS 7 control and is the answer not in the manual?

Table of Contents