For easy access to the information contained herein, keep this booklet at hand and review it as needed. Ask your local authorized Belmont dealer to install this product by following the installation instructions contained in the product. If you have any questions regarding the instructions for use or this product, contact your local authorized Belmont dealer.
Table of Contents Dear Customers ....................2 General Information 1-1 Intended Purpose of the Product ..............6 1-2 Compliance with Directives ................. 6 1-3 Declaration of Conformity ................6 1-4 How to Dispose of the Device ..............7 1-5 Disposal of Residues ................... 7 1-6 Concerning the water used for dental units ..........7 1-7 Recomendation to User ................
Page 4
Table of Contents Product Specifications 4-2 Overview and Major Components ............40 4-2-1 Over the Patient ..............40 4-2-2 Cart ..................41 Operation 5-1 Doctor unit section ..................42 5-1-1 Holder type ................42 5-1-2 Rod type ................46 5-1-3 Function swicth setup procedure .......... 50 5-1-4 Control knobs ..............
Page 5
Table of Contents Maintenance and Cleaning 6-1 Exterior ....................67 6-2 Doctor unit ..................67 6-3 Cuspidor unit ..................69 6-4 Suction line ..................70 6-5 High-volume evacuator hose/Saliva ejector hose ......... 70 6-6 Water line ..................... 70 6-7 High-volume evacuator/Saliva ejector ........... 71 77 type 3WAY syringe ................73 SYR-20 3WAY syringe ................
Product Type : DENTAL UNIT (CLASS II a) Product Name : CELEB " CELEB " has been defined by the rule 9 of MDR Annex VIII. The product has been designed and manufactured in accor- dance with the European standards as the listed in the Declara-...
1 General Information How to Dispose of the Device When disposing of this product and parts replaced, carefully take infection control measures and handle them properly in accor- dance with the relevant laws and regulations (applicable regula- tions and local regulations). In the EU area, EU Directive 2012/19/EU (Directive on Waste Electrical and Electronic Equipment [WEEE Directive]) applies to this product.
fit into some handpieces due to the manufacturing tolerances. Have your local authorized Belmont dealer to check the connect- ability before purchasing the handpiece. Except for our recommended handpieces, we shall not be liable for any problems deriving from bad connectability or their perfor- mance.
1 General Information 1-12 Symbols The symbols listed below are used on this product, on labeling, and in this booklet. Check the meaning of each symbol. Switch Switch (ON) (OFF) Switch for raising Switch for lowering the chair the chair Switch for reclining Switch for raising the backrest...
Page 10
1 General Information Electric motor speed Micromotor rotation set switch switch for nomal/reverese Start switch Coolant water ON/OFF Handpiece light switch switch (on/off) Function switch Scaler power control Power indicator Chair lock indicator Scaler mode selection E (ENDO) E S P 3 way syringe S (SCALING) P (PERIO)
1 General Information Classification of Type B Applied Parts foot controller An autoclave symbol that DRAIN indicates parts can be Drain valve VALVE sterilized in an autoclave at temperatures up to 135°C Dental unit Dental patient chair Manufacturing date and Name and address country of the manufacturer Third-party certification...
2 Safety Consideration 2-1 Risk Level Interpretation Precautions before use Make sure to carefully read the Safety Precautions and Operating Precautions and use the product correctly. These precautions are intended to ensure the safe use of the product and prevent harm or damage to users or other people. According to the magnitude of harm and damage and the degree of urgency, an incident that may be caused by misuse of the product is classified into one of the following categories: CONTRAIN-...
Page 13
2 Safety Consideration Classification by degree of harm or damage and urgency CONTRAINDICATION Use of the product without regard to this indication will create a hazardous condition that may result in death or serious injury. WARNING Improper handling of the product without regard to this indication will create a hazardous condition that may result in death or serious injury.
WARNING Installing or transferring Precautions for installation Ask your local authorized Belmont dealer to install the product. the product Make sure to place the product on a firm and flat floor. Placing the equipment on a non-flat floor may cause it to fall.
Page 15
2 Safety Consideration WARNING Keep moving parts of the product away from patients’ and Use and maintenance users’ hands, fingers, and bodies. of the product Hands, fingers or bodies may become caught in the product, causing injury. Do not wash the product with water. This may cause failure or electric shock.
Page 16
Discontinue use of the product if it is broken. In the case of a broken product, immediately discontinue use and turn off the main switch. Then, ask your local authorized Belmont dealer to repair the product. Turn off the main switch during cleaning.
Page 17
2 Safety Consideration WARNING Be sure to turn off the main switch upon completion of Use and maintenance work or during work breaks of the product Be sure to turn off the main switch upon completion of work or during work breaks. This prevents incorrect operation due to accidental contact and associated hazards.
Page 18
If you feel something is wrong, immediately discontinue use of the product and turn off the main switch. Then, contact your local authorized Belmont dealer. Confirm safety before use Before use, confirm that the parts are correctly and safely operating and that there are no obstacles around this product.
Page 19
Precautions when using a handpiece In the event of heat or smoky smell, stop using the handpiece, turn off the main switch, and contact your local authorized Belmont dealer. This may cause burns or have an adverse effect on the dental pulp.
Page 20
2 Safety Consideration CAUTION Use and maintenance Put on the cleaning gloves when detaching cuspidor bowl. of the product Cuspidor bowl is made of ceramic or glass that might cause injury when broken. Precautions when using water other than tap water The water unit is intended for use with tap water,purified water, distilled water, or pure water.
Page 21
2 Safety Consideration CAUTION Safety precautions regarding Conduct flushing before treatment. water quality If this product is not used for a long time, water retained in ducts and in the water heater will be susceptible to bacterial growth. To provide safe treatment and operate the handpiece without any trouble, never forget to perform flushing (discharge) of the water lines before starting treatment.
Page 22
2 Safety Consideration Preparation for flushing How to detach the handpieces Micromotor Ultrasonic Scaler Detaching position Detaching position Contra-angle handpiece Motor Hose Handpiece Hose Air Scaler Detaching position Straight handpiece Motor Hose Air Motor Detaching position Handpiece Coupling Hose Air Turbine Contra-angle handpiece Air motor Hose...
2 Safety Consideration Points to remember when operating the product Over the patient Meaning of a symbol Point to remember (locations requiring attention including moving parts, rotating parts, and detachable parts) Point to remember where an emergency stop mechanism is available Meanings of each point to remember ① Check the locking mechanism of the Cuspidor bowl.
DISTURBANCES is high. 4. Electromagnetic emission declaration Guidance and manufacturer’s declaration—electromagnetic emissions CELEB is intended for use in the electromagnetic environment specified below. The customer or user of CELEB should ensure that it is used in such an environment. Compliance Emissions test Electromagnetic environment—guidance...
Page 25
5. Electromagnetic immunity declaration 1 Guidance and manufacturer’s declaration—electromagnetic immunity CELEB is intended for use in the electromagnetic environment specified below.The customer or user of CELEB should ensure that it is used in such an environment. IEC 60601 Electromagnetic environment...
Page 26
6. Electromagnetic immunity declaration 2 Guidance and manufacturer’s declaration—electromagnetic immunity CELEB is intended for use in the electromagnetic environment specified below.The customer or user of CELEB should ensure that it is used in such an environment. IEC 60601 Electromagnetic environment...
Page 27
2 Safety Consideration Near electromagnetic field caused by RF wireless communication devices Test frequency Modulation IEC 60601 IEC 60601 test level compliance level (MHz) Pulse modulation 27V/m 27V/m 18Hz Frequency modulation 28V/m 28V/m ±5kHz shift 1kHz sine wave Pulse modulation 9V/m 9V/m 217Hz...
2 Safety Consideration Compatible Handpieces The following handpieces are compatible with this product: SYR-20 77 type 3Way DCI 3Way Syringe LUZZANI Minilight 3Way LUZZANI Minilight 6Way NSK Ti-Max Z Series NSK Ti-Max X Series Air turbine NSK S-Max M Series * Not optic turbine NSK Pana Max Plus Series for A-type...
2 Safety Consideration Compatible Dental Chairs The following Dental Chairs are compatible with this product: CELEB CHAIR (EURUS TYPE) Dental Chair CELEB CHAIR Compatible Dental Lights The following dental lights are compatible with this product: EURUS LIGHT Dental Light 900 Dental Light (Type 920/Type D200)
In the case of malfunction and contact information In the case of any problems, discontinue use, turn off the main switch and contact your local authorized Belmont dealer. Action to take in the case of a water leak In the event of a water leak, close the main water valve, turn off the main switch and breaker for devices used in the clinic, and contact your local authorized Belmont dealer.
Abbreviation Chair Mount Over the Patient Type CM Chair Mount Cart Type CM CT Pedestal Over the Patient Type Pedestal Cart Type PD CT CM CT PD CT CELEB CHAIR Dental chair ○ ○ ○ ○ (EURUS TYPE) ○ ○...
Size: 5.2 × 20 mm 99.5Kg (without dental light) Weight Doctor table 3 kg Weight limit Sub-tray of doctor table (Rod) 1.5 kg CELEB CHAIR (EURUS TYPE) (maximum patient mass;) 200 kg CELEB CHAIR (maximum patient mass;) 150 kg Air supply Main air pressure 0.5MPa...
Page 33
The product is not suitable for use in a environment high-oxygen environment. Dimensional drawing (standard values are provided) 340˚ 290˚ ø50.8 (ø42.7) 21˚ 28˚ CELEB CHAIR CELEB CHAIR (EURUS TYPE) 900/300 EURUS Dental Light Dental Light LIGHT Unit: mm 1479 1554 Tolerance in dimensions: ±10%...
Operating speed: Time lag Size: 5.2 × 20 mm 99.5Kg (without dental light) Weight Doctor table 3 kg Weight limit CELEB CHAIR (EURUS TYPE) (maximum patient mass;) 200 kg CELEB CHAIR (maximum patient mass;) 150 kg Air supply Main air pressure 0.5MPa...
Page 35
Atmospheric pressure 700 to 1060 hPa Adaptation to high-oxygen The product is not suitable for use in a environment high-oxygen environment. Dimensional drawing (standard values are provided) ø50.8 (ø42.7) CELEB CHAIR CELEB CHAIR (EURUS TYPE) Unit: mm 900/300 EURUS Dental Light Tolerance in dimensions: ±10%...
Size: 5.2 × 20 mm 111.7Kg (without dental light) Weight Doctor table 3 kg Weight limit Sub-tray of doctor table (Rod) 1.5 kg CELEB CHAIR (EURUS TYPE) (maximum patient mass;) 200 kg CELEB CHAIR (maximum patient mass;) 150 kg Air supply Main air pressure 0.5MPa...
Page 37
4 Product Specifications Usage environment Temperature 0°C to 40°C Humidity 10% to 95% (No condensation) Atmospheric pressure 700 to 1060 hPa Transportation/storage Temperature −20°C to 70°C environment Humidity 10% to 95% (No condensation) Atmospheric pressure 700 to 1060 hPa Adaptation to high-oxygen The product is not suitable for use in a environment high-oxygen environment.
Operating speed: Time lag Size: 5.2 × 20 mm 111.7Kg (without dental light) Weight Doctor table 3 kg Weight limit CELEB CHAIR (EURUS TYPE) (maximum patient mass;) 200 kg Air supply Main air pressure 0.5MPa Filter mesh size 50 m Minimum flow rate...
Page 39
4 Product Specifications Usage environment Temperature 0°C to 40°C Humidity 10% to 95% (No condensation) Atmospheric pressure 700 to 1060 hPa Transportation/storage Temperature −20°C to 70°C environment Humidity 10% to 95% (No condensation) Atmospheric pressure 700 to 1060 hPa Adaptation to high-oxygen The product is not suitable for use in a environment high-oxygen environment.
Junction unit Built-in chair type - CELEB CHAIR - Cuspidor unit Junction cover Close Drain valve Open knob Doctor unit - CELEB CHAIR (EURUS TYPE) Junction cover Close Open External type Utility box Holder for intra-oral Camera Drain valve knob...
5 Operation Doctor unit section 5-1-1 Holder type Master swith Master Switch Turn on the master switch located under the right hand side of the doctor table, the power indicator on the main control panel will then illuminate green. Turn off the master switch after daily operation and for long term interval.
Page 43
While the bowl flush is working by momentarily pressing the bowl flush switch ( ) , the bowl flush will stop. Note : CELEB can be set to timer mode (standard setting) and continuous mode for bowl flush. The bowl flush water volume can be adjusted by the bowl flush flow control knob located inside the cuspidor body.
Page 44
5 Operation Chair lock indicator The safety lock indicator illuminates umber when the safety lock device is working. Note: Please refer to 5-5. Lock function Coolant water ON/OFF switch When a handpiece is picked up and this switch is pressed, both LED A (air) and LED W (water) lights up, the coolant water and air comes out from the handpiece.
Page 45
5 Operation Micromotor rotation switch for nomal/reverese Amber LED Green LED (Reverse rotaition) (Nomal rotaition) After picking up the electric motor from the holder, the electric motor rotation direction can be changed by momentarily pressing this switch ( ) the rotation direction will be indicated by the amber and green LEDs.
5 Operation 5-1-2 Rod type Master swith Master Switch Turn on the master switch located on the right hand side (facing) under the doctor table,the power indicator on the main control panel illuminates in green. Main control panel A Spec. control panel ①...
Page 47
While the bowl flush is working by momentarily pressing the bowl flush switch ( ) , the bowl flush will stop. Note : CELEB can be set to timer mode (standard setting) and continuous mode for bowl flush. The bowl flush water volume can be adjusted by the bowl flush flow control knob located inside the cuspidor body.
Page 48
5 Operation Chair lock indicator The safety lock indicator illuminates umber when the safety lock device is working. Note: Please refer to 5-4. Lock function Coolant water ON/OFF switch When a handpiece is picked up and this switch is pressed, both LED A (air) and LED W (water) lights up, the coolant water and air comes out from the handpiece.
Page 49
5 Operation Micromotor rotation switch for nomal/reverese Amber LED Green LED (Reverse rotaition) (Nomal rotaition) After picking up the electric motor from the holder, the electric motor rotation direction can be changed by momentarily pressing this switch ( ) the rotation direction will be indicated by the amber and green LEDs.
5 Operation 5-1-3 Function swicth setup procedure 1. Timer TIMER SET START Timer can be set maximum 90 mins. 50 secs. in 10 secs. segment. Function Switch Decrease Switch Increase Switch Start switch 1) To set a timer Momentarily press the function switch, and set the time by pressing decrease switch and increase switch.
Page 51
FLUSH OUT 3. Flush out system The CELEB is equipped with two types of flush out system. Short time flush out is for cleaning handpiece water lines. Long time flush out is for handpiece water lines, bowl flush water line and cupfiller water line.
Page 52
5 Operation 4. Control panel switching sound on/off Beep Pressing a switch on the control panel makes an electronic sound. This sound can be eliminated as follows; Function Switch Decrease Switch Increase Switch Momentarily press the function switch four times and momentarily press the decrease switch.
Page 53
5 Operation 8. Coolant water ON/OFF switch SPRAY MODE In case of micromotor, either the 2-mode or the 4-mode can be MODE selected by mode select setup. When this switch is pressed in the 2-mode setup, switching between spray and OFF occurs. In case of 4-mode setup, switching occurs in the sequence indicated below each time when this switch is pressed: Spray to Water only to Air only to OFF.
5 Operation 5-1-4 Control knobs Holder type Handpiece spray water control knobs The handpiece spray water control knobs are located under the doctor table. Each handpiece spray water control knob is marked 1-4 from the left side HP1,HP2,HP3,... The handpiece spray water volume can be controlled independently.
5 Operation 5-1-6 Balance arm air brake Holder type Balance arm Balance arm air brake button is located on the handle. When the master switch is ON, the balance arm is locked. Grasp the handle and press the air brake button to adjust the table height.
5 Operation 5-1-9 Handpieces The handpiece is actuated by picking it up from the handpiece holder and operating the foot controller. Operation of the each handpieces, please refer to the manufacturer’s instruction manual attached to the individual equipment. CAUTION Proper Improper Repeated actions of picking up and returning the handpiece may cause its hose twisted.
5 Operation 5-1-11 Intraoral camera holder bracket Intraoral camera holder can be mounted on this bracket. [Note] Mountable intraoral camera holders: Holder for SOPRO ACTEON SOPRO 617 Intraoral camera holder bracket 5-1-12 Sub tray holder WARNING A load exceeding the weight limit (1.5 kg) must not be applied to the sub-tray of the doctor table (Rod).
5 Operation Cuspidor unit section 5-2-1 Cuspidor unit control panel Dental light mode selection switch MANUAL SENSOR EURUS LIGHT 900 Dental Light Water heater switch Water flow control knob for service coupler for water use Cuspidor unit Service coupler for water use Service coupler for air use control panel Dental light mode selection switch...
While the bowl flush is working by momentarily pressing the bowl flush switch ( ) , the bowl flush will stop. Note : CELEB can be set to timer mode (standard setting) and continuous mode for bowl flush. The bowl flush water volume can be adjusted by the bowl flush flow control knob located inside the cuspidor body.
5 Operation Chair auto mode switches Preset operation Momentarily press the preset-I switch ( the chair moves to the preset 1 position and stops automatically. Preset 2 position operated by the preset switch ( Note : For preset position adjustment refer to chair Manual. Automatic return operation Momentarily press the auto return switch ( the chair returns to the initial position (the seat is the lowest...
5 Operation 5-2-4 VH-18 HVE (High-volume evacuator) HVE Tip Take the HVE out of the assistant holder, and suction will start. Since a delay circuit is provided, suction does not stop immediately when the HVE is returned to the assistant holder. Open In case of the central suction system, suction will continue for approx.
5 Operation 5-2-8 Clean water system Pressure control knob CITY BOTTLE select switch can be changed between municipal Pressurization ON-OFF switch water and water bottle. CITY BOTTLE CITY..Tap water select switch BOTTLE... Water bottle When the pressurization ON-OFF switch is set to ON, the water bottle may be used.
5 Operation 5-2-10 Dental light Before use, always read the Instructions for dental light to ensure correct usage. 300 (Type320S) EURUS LIGHT 900 (Type D200) 900 (Type 920) 5-2-11 Monitor Bracket the PC monitor can be mounted on this bracket. CAUTION Monitor bracket Do not apply an excessive load or shock to the monitor or...
5 Operation 3WAY Syringe SYR-20 Angle Type Straight Type Spraying water / air Press W lever to have water come out. Press A lever to have air come out. Press both levers simultaneously to have spray come out. Water comes out from the center of the tip, and air comes out from the circumference of the tip.
5 Operation 10˚ Foot controller Rod type Pick up a handpiece from the handpiece holder or the handpiece rest (pull the rod about 10 degrees to forward). SE Type Handpiece Rest Pedal SE Type Pick up a handpiece from the handpiece holder and depress Coolant water the drive air pedal, the handpiece starts running.
5 Operation 5-5 Lock function In the following cases the safety lock device to lock the chair Chair lock indicator movement is activated. 1. When the pedal of the foot controller is depressed. 2. Press any chair operation switches for stopping the automatic movement.
7 Maintenance and Cleaning by Dental Staff 6 Maintenance and Cleaning 6-1 Exterior Cleaning and disinfecting the surfaces To clean and disinfect the exterior of the product, wipe the surface with a soft cloth or paper towel moistened with FD366 manufactured by Dürr / PlastiSept eco Wipes FP manufactured by ALPRO, and then wipe it with a dry cloth.
Page 68
If the hygroscopic sponge (2) (consumable) is excessively dirty (1) Oil reservoir or has excessive oil on it, replace it. Contact your local authorized Belmont dealer for a replacement line for the hygroscopic sponge. (Replacement will be charged for.) Handpieces/Handpiece hose Cleaning of handpieces Refer to the Instructions for Use for the respective handpieces.
7 Maintenance and Cleaning by Dental Staff 6 Maintenance and Cleaning 6-3 Cuspidor unit Cupfiller Base Drain Cap, Basket Strainer, Cupfiller base, Cuspidor Bowl Drain Cap A. Detach the drain cap, and clean the basket strainer. Remove the cupfiller base and clean it. Turn the bowl flush nozzle counterclockwise, avoiding the clash Basket with the cuspidor bowl.
7 Maintenance and Cleaning by Dental Staff 6 Maintenance and Cleaning 6-4 Suction line Handpieces for HVE and saliva ejector suction the secretions, saliva, or blood that contains bacteria. Therefore, always wash and sterilize them using MD555/Orotol plus manufactured by Dürr after the procedure. Use Orotol Plus for daily care.
7 Maintenance and Cleaning by Dental Staff 6 Maintenance and Cleaning 6-7 HVE and saliva ejector (High-volume evacuator) (1) Disassembly Wash and sterilize the handpieces between patients. To properly sterilize the HVE and saliva ejector, it is necessary saliva ejector to wash them to remove dirt and immerse them in detergent.
Page 72
7 Maintenance and Cleaning by Dental Staff 6 Maintenance and Cleaning (5) Sterilization The HVE and saliva ejector are autoclavable. Caution when assembling Reassemble the main body unit, and autoclave the HVE and HVE before sterilization saliva ejector. When assembling the HVE, align the two surfaces; a flat surface of the cap and the surface of the sliding groove of the body.
7 Maintenance and Cleaning by Dental Staff 6 Maintenance and Cleaning 6-8 77 type 3WAY syringe Wash and sterilize the handpieces between patients. To properly sterilize the 3way syringe, it is necessary to wash them to remove dirt and immerse them in detergent. Then, rinse them to remove any remaining detergent.
Page 74
7 Maintenance and Cleaning by Dental Staff 6 Maintenance and Cleaning (3) Sterilization (3) Sterilization Choose the appropriate method from the following sterilization cycles depending on the type of the autoclave sterilizer in your clinic: [Dynamic-Air-Removal] 1. Put the nozzle in a sterilization bag, and seal the opening. 2.
7 Maintenance and Cleaning by Dental Staff 6 Maintenance and Cleaning 6-9 SYR-20 3WAY syringe Cleaning the inside of the lever cotton swab Lever If dust or dirt accumulates inside the lever, use a cotton swab to remove them. Wash and sterilization Wash and sterilize the handpieces between patients.
Page 76
7 Maintenance and Cleaning by Dental Staff 6 Maintenance and Cleaning (3) Sterilization (3) Sterilization Choose the appropriate method from the following sterilization cycles depending on the type of the autoclave sterilizer in your clinic: [Dynamic-Air-Removal] 1. Put the nozzle in a sterilization bag, and seal the opening. 2.
The water filter in the junction box needs to be replaced at least once a year. The air filter in the junction box needs to be replaced at least once every three years. Contact your local authorized Belmont dealer. Air filter Water filter CELEB Chair (EURUS TYPE) Air filter...
The chair unexpectedly If the chair does not stop, please contact Make sure the chair movement Check the Before moves during procedure, your local authorized Belmont dealer. cancel consulting stops by any of the following resulting in an injury. The function hours actions.
Page 79
[Refer:6 Maintenance and Cleaning ] Saliva ejector hours Check the Before Water, air, and spray come out The product does not Contact your local authorized Belmont function properly, functioning consulting by pressing A lever and W lever. dealer. preventing proper...
Maintenance and inspection can be outsourced to qualified persons such as authorized repairers of medical devices. If you have any question about periodic inspection, contact your local authorized Belmont dealer. List of service parts required for the periodic inspection Standard...
6 Maintenance and Cleaning 6-14 Detachable parts Part name Part name Handpiece HVE tip Handpiece hose Solid collector lid Cuspidor bowl Solid collector filter Drain cap Water service coupler Basket strainer Air service coupler Cupfiller base Syringe nozzle Cupfiller nozzle Syringe body Oil mist separator Saliva ejector...
Refer to ‘Troubleshooting’ before you check the device. If the problem persists, turn off the main switch, and contact your local authorized Belmont dealer to request a repair. Service life The service life of this product is 10 years on condition that maintenance and inspection are properly conducted [according to our self-certification (our data)].
If you encounter any of the problems listed below, take the countermeasures described below before requesting a repair. If the problem persists even after troubleshooting, stop using the product immediately, turn off the main switch, and contact your local authorized Belmont dealer. Phenomenon Please check Remedy...
9 Consumables Consumables are parts that will normally wear or deteriorate, change their appearance, or become damaged after use. Please note that repair or replacement of consumables are not covered by the warranty and will be charged for. (* Degree of wear, deterioration or damage and timing for replacement depends on the use environment and conditions at the customer’s premises.) Consumables (Parts listed below are out of the guarantee...
Page 85
TAKARA COMPANY EUROPE GmbH Berner Strasse 18, 60437 Frankfurt am Main, Germany TEL : +49-69-506878-0 FAX : +49-69-506878-20 0197 TAKARA BELMONT CORPORATION 2-1-1, Higashishinsaibashi,Chuo-ku,Osaka, 542-0083, Japan TEL : +81-6-6213-5945 FAX : +81-6-6212-3680...
Need help?