Omron Z5FP-MC10 Setup Manual page 27

Multi-dimensional sensor
Table of Contents

Advertisement

Measurement range
Reference distance
Measurement
center
Z550-SW70
(Installation is possible for diffuse reflection only. Installation for regular (mirror) reflection is not
possible.)
Measurement
Reference distance
range
(Diffuse reflection)
10mm
178mm
30mm
192mm
60mm
210mm
100mm
230mm
160mm
260mm
240mm
300mm
Select a mounting condition of the sensor in [System/Environment] from the menu, and then mount
the sensor at a distance where it can accurately measure workpieces.
If the connector chassis of a sensor or extension cable makes contact with a noise source, operation
may not proceed normally. Insulate the connector chassis before use.
Insulation is necessary.
Sensor
Sensor
H
Measurement range
(Height)
L
Measurement range
(Width)
Measurement
range
Direction of height
10mm
62mm
30mm
63mm
60mm
64mm
100mm
64mm
160mm
64mm
240mm
64mm
Extension Cable
Controller
SECTION 2
Installation
Measurement
Measurement
range
range
Width direction
Width direction
(H side)
(L side)
65mm
72mm
81mm
92mm
109mm
132mm
POWER
25
Setup Manual