2. Advanced functions

When activated in the configuration menu, all advanced functions can be reached by pressing
Then, press
to reach required item and
Pressure
Total volume
Bat.autonomy
Note: Upon pump status (stop or infusion) certain functions may not be available, example : Loading dose not available after infusion start.
2.1.
Pressure monitoring and setting
1.
Press
, then
.
 Line disconnection / Drop in pressure alarm
P
ON
P
ON
 Dynamic Pressure System
Pressure increase control
2.
Press
to select item.
3.
Press
to select value.
4.
Return to main screen by pressing
2.2.
Total volume / Infused volume by
mode
1.
Press
.
Pressure
Total volume
Bat.autonomy
2.
Press
to select "Total Volume".
3.
Using the
key, the infused volumes of the different modes
(Secondary, Primary, Ramp up / down, Sequential) are displayed.
4.
To clear the total infused volume, the infusion should be on hold.
Switch to the total volume screen and use the
the
key to restore).
When total volume is cleared, all infused volumes of the different
modes are also cleared.
PRI
VP total=
5.
Confirm by pressing
- 4 -
Loading dose
Ramp up / Down
Sequential
/
700
 Pressure alarm threshold
mmHg
 Infusion pressure
50
.
key to clear (use
0100 ml
00100 ml
in
2h00
. Then, press
to escape.
to enter desired screen.
Bolus
Secondary
Pause
/
2.3.
The present battery charge level can be reached by pressing either
or
1h30 / 120 ml/h
See paragraph 11 for functioning with battery.
2.4.
1.
Access with the
2.
If necessary, select the volume of bolus to infuse with
3.
Select the bolus duration with
automatically calculated.
4.
Select the flow rate bolus with
automatically calculated and readjusted according to displayed flow
rate.
5.
Check bolus parameters and start bolus infusion by pressing
6.
After the bolus has been delivered, a short beep is emitted. The
pump automatically resumes its initial status (stand-by or infusion)
and the bolus programming parameters are stored. If the bolus is
interrupted, the remaining bolus volume to infuse is kept in memory.
Note that bolus infused volume is subtracted from remaining volume to
infuse.
Note: for a bolus at infusion start, use Loading dose mode.
.
Bolus
Secondary
Keys locked
/
Battery charge level
:
 Average battery life at current
flow rate
90%
 battery charge indication
Bolus
keys.
10ml
1min0
Total IV
600ml/h
0212ml
. Then, confirm and shift to next item with
1310-4_nu_module_mvp_st_eng_140128.doc
.
. The bolus rate is
. The bolus duration is
.