What to do if...
If alarm occurs:
Mute alarm (if necessary)
Detect the cause and proceed to the appropriate action
Release alarm
Resume feeding
Empty battery
Connect pump to
mains supply
immediately.
Door open alarm
Close pump door.
Set installation alarm
Check giving set
position above and
below pump
mechanism and
insert it correctly
FOR ANY
ENQUIRIES
OR INFORMATION CALL THE HOMECARE
24 HOUR FREEPHONE HELPLINE NUMBER 0808 100 1990
Fresenius Kabi Limited,
Cestrian Court, Eastgate Way,
Manor Park, Runcorn,
Cheshire WA7 1NT
caring
for life
tel: 0808 100 1990
www.fresenius-kabi.com/gb
Date of preparation: October 2023 GB-Ami-2300001
Start Reminder
Proceed to next step
or switch pump off.
Empty bag/ air in
Downstream
set alarm
occlusion alarm
Change feed
Check that pumpset/
container, check
feeding tube is clear
connection to
of any occlusion after
giving set and
the pump, at the
ensure giving set is
patient side.
fully primed.
Upstream
Target volume
occlusion alarm
almost reached
message
Check that pump
set is clear of any
Continue feeding
occlusions between
or end feeding.
the container and
the pump
Target volume
Pump error alarm
reached alarm
Note the technical
End feeding or
Error code (Err xyz)
proceed to the
and contact your
next step
biomedical
department.
Amika
®
is a registered trademark of Fresenius Kabi AG.
Fresenius Kabi Ltd is an authorised user © Fresenius Kabi Ltd.
caring for life
Amika
®
ENTERAL NUTRITION PUMP
PICTORIAL REFERENCE GUIDE