Technical specifications

10.1

Power

Table 10-1
Description
Power supply
Battery voltage
1
Only use the standard lithium battery pack RRC2054 as supplied by Siemens.
Only use battery station (RRC-SMB-MBC) and power supply PS65 (A11-065N1A model) or equivalent as
2
supplied by Siemens.
10.2

Operating conditions

Table 10-2
Description
Ambient temperature
(Humidity max. 90 %)
Degree of protection while connected to flowmeter
Environmental class
Altitude
Pollution
Overvoltage category
EMC performance
Verification accuracy
Recalibration and traceability Calibration at Siemens Hagenau (COFRAC accredited laboratories) trace‐
1
With connector lid open and no cable plugged in, the degree of protection is IP20.
SITRANS Verificator
Operating Instructions, 03/2024, A5E51481698-AA
Power supply
Basic conditions
Operation:
Storage (with battery):
Storage (without battery):
Charging battery:
Emission
Immunity
Fingerprint
Insulation threshold
Coil resistance
(excluding temperature variations)
Transmitter test
able to national and international standards. Mandatory annual recali‐
bration to maintain traceability
Specification
Removable battery pack
1
14.4 V DC
2
Specification
-20 °C to +50 °C (-4 °F to +122 °F)
-20 °C to +50 °C (-4 °F to +158 °F)
-40 °C to +70 °C (-40 °F to +158 °F)
0 °C to 45 °C (+32 °F to +113 °F)
IP54
1
4M11
Up to 2000 m (6560 ft)
Degree 3
II
EN 300328
IEC 61326-1
2 %
2.5 mm/s
1 %
1.2 %
10
45