REGULAR CHECKS DURING SAMPLING
12
Do not exceed analyser operating specifications:
Do not exceed probe maximum temperature
•
Do not exceed analyser internal temperature
•
Do not place analyser on a hot surface
•
Keep analyser water trap vertical - water vapour condenses in probe line
•
and can quickly fill analyser water trap
Keep analyser in-line particle filter clean and dry
•
NORMAL SHUTDOWN SEQUENCE
13
DO THIS EVERY TIME YOU USE THE ANALYSER
Remove probe from exhaust - TAKE CARE! HOT PROBE
- and allow to cool naturally.
Allow analyser to purge in fresh air for at least three minutes or until all toxic
sensor readings are below 10ppm.
Do not immerse probe in water as this will damage pump and sensors.
NOTE: It is good practice to hang probe hose vertically after sampling so
condensate drains away.
KANE-EGA4/5 MANUAL 29