Specifications

INPUT POWER
Voltage: 2.8 V
DIO DIGITAL INPUTS
V
inHighThreshold
V
inLowThreshold
V
= 30 V
inMAX
DIO DIGITAL OUTPUTS
2x Open Drain 100 mA max
V
= 30 V
MAX
ACCURACY
Temperature
Range: -40 to 85°C (-40 to 185°F)
Accuracy at 25°C: ±0.3°C (±0.6°F)
Response Time: Less than 1 second
Temperature Coefficient: less than 0.01 C/C
Repeatability: ±0.15°C
Relative Humidity
Accuracy at 25°C: ±2.5% (0 to 80%), non-condensing ±3.5% (80 to 100%), non-condensing
Hysteresis: ±0.8%
Response Time: 8 seconds
Repeatability: ±0.21% RH
Barometric Pressure
Accuracy Over Full Range: ±6 mbar from
Accuracy @ 25°C: ±4 mbar from 700 to 1100 mbar
Dewpoint (Calculated)
Accuracy: ±2°C
Heat Index (Calculated)
Accuracy: ±2°C
Humidex (Calculated)
Accuracy: ±2°C
ENVIRONMENTAL
Operating Temperature: -40 to 85°C (-40 to 185°F)
MECHANICAL
Tube Housing
Construction: Stainless Steel
Dimensions: 72 mm L x 15.9 mm OD (2.83" x 0.62") (5" model available)
Bulkhead Housing
Construction: Aluminum
Dimensions: 73 mm L x 15.9 mm OD x 18.5 mm Panel Opening (2.83" x 0.62" x 0.72")
GENERAL
Agency Approvals: CE, UKCA
Configuration: Configurable via Omega Link Smart Interface and SYNC configuration software
Software: Compatible with OEG, SYNC, and OMEGA Cloud
- 3.3 V
DC
DC
= 2.2 V
MAX
= 0.3 V
MIN
DC
DC
300 to 1100 mbar
8 |