Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19022/19022-4
Quick Start Guide
2.
The cable quantity is varied with the output channel configured,
for instance, the cable quantity for 10CH model is 10 sets and for
4CH mode is 4 sets.

1.3 Inspection

Before shipment, this instrument was inspected and found to be free of
mechanical and electrical defects. As soon as the instrument is unpacked,
inspect for any damage that may have occurred in transit. Save all packing
materials in case the instrument has to be returned. If damage is found, please
file claim with carrier immediately. Do not return the instrument to Chroma
without prior approval.
2.

Precaution before Use

The Hipot Tester can output up to 8KV high voltage for external test. Accidents
may occur or even cause death if using this Tester incorrectly or in the wrong way.
Thus for safety sake, be sure to read the precautions in this chapter to avoid any
accidents from happening.
High voltage module replacement
If users replace or switch the 10 sets of high voltage modules on the rear
panel, to ensure the Tester output is still within the specification, users need
to recalibrate the Tester and ensure the specification.
Electric shock
To prevent the incident of electric shock from occurring, it is suggested to
wear the insulated rubber gloves before using the Hipot Tester for electricity
related tasks.
Grounding
A safety ground terminal is located at the rear of the Tester chassis; please
use a proper tool to ensure it is grounded accurately. If not it would be very
dangerous when the power circuit or the connection cable of any device
shorts with ground terminal as the chassis may contain high voltage.
Anyone who touches the device in this case may cause electric shock.
Therefore, it is necessary to connect the safety ground terminal to earth
properly as the arrow shows in section 3.2.
Connecting test cable to
It is necessary to check if the test cable is connected all the time when the
Tester is in use. When connecting a test unit with test cable, connect the
RTN/LOW test cable to the UUT first. When the host RTN/LOW terminal is
connected, it is very dangerous if the test cable on RTN/LOW terminal is not
2
terminal
RTN/LOW