Download Print this page

Assemble And Check Fit-Up - Emerson Rosemount 485 Annubar Flanged Quick Start Manual

Advertisement

March 2024
3 

Assemble and check fit-up

For accurate measurement, use the following steps to ensure that
Ports A and B are equal distances from the inside walls of the pipe.
Procedure
1. Assemble the Rosemount 485 to the mounting hardware with
the gaskets and bolts.
2. Hand-tighten the bolts just enough to hold the position of the
sensor centered in the mounting hardware.
3. Measure the distance from the high point of the butt weld
branch connection to the first sensing hole, port B, then
subtract 1/16 in. (1.6 mm).
4. Measure the distance from the end of the transferred length
in
5. Compare the numbers obtained in
Note
Small discrepancies can be compensated for with the fit-up
of the mounting hardware. Large discrepancies may cause
installation problems or error.
Quick Start Guide
Step 3
to the last sensing hole, port A.
Quick Start Guide
Step 3
and Step 4.
11

Advertisement

loading