12-1-29. Read Scale Function Settings

This command is supported only in ADP18 or its later versions, plus AFP15 or its later versions.
Write UUID E812 handle (Hex) with the following command(s):
Calibration Command (Hex)
99
After the above 'Read' command set is successfully written to the UUID E812 handle, there will be
return values (Hex data) after reading the UUID E812 handle, such as the following:
99 03 0F ZZ ZZ (for descriptions regarding the ZZ return value(s), please refer to 12-1-28. Write Scale
Function Settings)
If the 'Write' operation has failed, reading UUID E812 handle will instead return 99 02 7F .... (please
refer to 12-1-39 Return Read & Write Status).
Execution conditions for these
commands

12-1-30. Write Bluetooth Mode

This command is supported only in ADP18 or its later versions, plus AFP15 or its later versions.
Write UUID E812 handle (Hex) with the following command(s):
Calibration
Length
Sub-Command
Command
(Hex)
(Hex)
(Hex)
99
02
50
N1: Calibration Mode Being Enabled status (this bit can only be read and cannot be written).
N1
Description
1
Calibration Mode Being Enabled
0
Normal Mode
H1: Connection Password Decryption Required for Every Connection
P1: Pairing required
H1
P1
Description
0
0
Allow any device to make a connection. Anyone can make a connection, but only one
connection is allowed at one time.
1
0
Allow a device with valid Connection Password to make a connection. Anyone with valid
Connection Password can make a connection, but only one connection is allowed at one
time.
0
1
Pairing must be completed once before making any connection. Only one previously paired
device is allowed to make a connection. Others cannot make a connection.
1
1
Pairing must be completed once before making any connection. Only one previously paired
device with valid Connection Password is allowed to make a connection. Others cannot
make a connection.
TM
Length (Hex)
01
Calibration Mode
Executable
1-byte Type A Data (Hex)
0 N1 H1 0 0 0 0 P1
52
Sub-Command (Hex)
0F
Normal Mode
Executable
ZSME000000026