Read Temporarily Switch E813 Notify Output Formats - Excell Wireless Platform Service Manual

Table of Contents

Advertisement

EXCELL PRECISION CO., LTD.

12-1-2. Read Temporarily Switch E813 Notify Output Formats

Write UUID E812 handle (Hex) with the following command(s):
Calibration Command (Hex)
99
After the above 'Read' command set is successfully written to the UUID E812 handle, there will be
return values (Hex data) after reading the UUID E812 handle, such as the following:
Calibration
Length
Sub-Command
Command
(Hex)
(Hex)
(Hex)
99
02
01
If the "Read UUID E812 handle" operation has failed, the return values will instead be 99 02 7F ....
(please refer to 12-1-39 Return Read & Write Status).
Execution conditions for these
commands
TM
Length (Hex)
01
1-byte
'Read' Description
Type A
Data (Hex)
AD Mode 1, Notify AD raw value (enable Notify, please
00
refer to 0x85 Commands)
AD Mode 2, Notify AD value after filtering (enable Notify,
01
please refer to 0x85 Commands)
AD Mode 3, Notify AD filtering -> value after linear
02
correction (enable Notify, please refer to 0x85
Commands)
Weight Mode 1 (enable Notify weight, traditional Excell
04
RS232 Formats ,"ST,NT,30.000Kg")
Weight Mode 2 (enable Notify weight, please refer to
05
0x84 Commands)
Weight Mode 3, power-on Notify weight, FOR
06
APP, "ST,N ,30.000Kg"
Calibration Mode
Executable
38
Sub-Command (Hex)
01
Normal Mode
Executable
ZSME000000026