Read Single-Point Weight Calibration; End The Calibration - Excell Wireless Platform Service Manual

Table of Contents

Advertisement

EXCELL PRECISION CO., LTD.
Read UUID E812 handle, which will return 99 02 7F .... (please refer to 12-1-39 Return Read & Write
Status).
Example :
For 10kg weight calibration, empty-load AD value reading after filtering is 2000; 10kg-load AD value
after filtering is 202000; 10kg will be displayed as 10.000kg. Therefore the weight display value
excluding the decimal points would 10000.
Therefore ZZ ZZ ZZ ZZ → 2000 converted to Type A Data → 00 00 0F 50
SS SS SS SS →202000 converted to Type A Data →00 0C 2A 10
WW WW WW WW →10000 converted to Type A Data →00 00 4E 10
Write UUID E812 handle (Hex) data 99 0D 48 00 00 0F 50 00 0C 2A 10 00 00 4E 10 . After successful
'Write' operation, you have now completed the single-point weight calibration.
Execution conditions for these
commands

12-1-16. Read Single-point Weight Calibration

Write UUID E812 handle (Hex) with the following command(s):
Calibration Command (Hex)
99
After the above 'Read' command set is successfully written to the UUID E812 handle, there will be
return values (Hex data) after reading the UUID E812 handle, such as the following:
99 0D 08 ZZ ZZ ZZ ZZ SS SS SS SS WW WW WW WW (for descriptions regarding the ZZ SS WW return
value(s), please refer to 12-1-15. Write Single-point Weight Calibration).
If the 'Write' operation has failed, reading UUID E812 handle will instead return 99 02 7F .... (please
refer to 12-1-39 Return Read & Write Status).
Execution conditions for these
commands

12-1-17. End the Calibration

This 49(Hex) sub-command set can be written only when Calibration Mode is enabled.
Write UUID E812 handle (Hex) with the following command(s):
Calibration
Length
Sub-Command
Command
(Hex)
(Hex)
(Hex)
99
02
49
Read UUID E812 handle, which will return 99 02 7F .... (please refer to 12-1-39 Return Read & Write
Status )
Execution conditions for these
commands
TM
Calibration Mode
Executable
Length (Hex)
01
Calibration Mode
Executable
1-byte Type A Data
'Write' Description
(Hex)
01
End the Calibration Mode
Calibration Mode
Executable
45
Normal Mode
Not executable
Sub-Command (Hex)
08
Normal Mode
Executable
Normal Mode
Not executable
ZSME000000026